Evaluation of Virtual Monoenergetic Images on Pulmonary Vasculature Using the Dual-Layer Detector-Based Spectral
Abed Ghandour, Andrew Sher, Negin Rassouli
1Philips Healthcare, Cleveland, OH.
Objective:
To evaluate the ability of retrospectively generated virtual monoenergetic images (VMIs) from the detector-based spectral computed tomography (SDCT) to augment pulmonary artery enhancement in CT and if iodine map can predict the optimal monoenergetic level.
Methods:
The study included 79 patients with contrast-enhanced chest CT scans on an SDCT scanner. Conventional 120-kVp images and VMI from 40 to 80 keV were generated. Attenuation, noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were measured at 7 different locations in the pulmonary arterial system. The iodine concentration (in milligrams per milliliter) was calculated using the iodine-density images. The overall image quality was subjectively graded on a 5-point scale, with 1 being the worst and 5 the best. Fifty-four patients with suboptimal pulmonary enhancement (<200 Hounsfield units [HU]) were then identified. From the VMIs, an ideal set was chosen that maintained mean vascular attenuation greater than 200 HU while maintaining at least diagnostically acceptable quality (ie, IQ score ≥3). At this ideal energy level, quantitative and qualitative parameters were compared with the standard 120-kVp polyenergetic study. Average iodine concentrations were correlated with the optimal keV levels used for salvaging suboptimal studies.
Results:
The mean attenuation of all the measured pulmonary arterial regions in the suboptimal cases was 136.1 ± 18.1 HU in conventional 120-kVp images. Attenuations of the VMIs at 40, 50, and 60 keV were significantly higher than conventional images measuring 357.5 ± 19.5, 243.6 ± 16.7, and 176.6 ± 15.0 HU, respectively (P < 0.001). Similar results were seen with SNR and CNR. In total, 50 studies can be salvaged, with 50 keV being the optimal energy for 21, 60 keV optimal for 17, and 40 keV optimal for 12 studies. At the optimal energy level, there were improvements of attenuation, SNR, and CNR by 71%, 63%, and 137% compared with conventional images. There was a positive correlation between iodine value and optimal reconstruction energy with a linear equation y = 5.9539x + 27.434 and R = 0.8093.
Conclusions:
Suboptimal enhanced pulmonary arterial CT studies can be salvaged using low-energy VMI generated from the SDCT scanner. There were significant improvements of attenuation, SNR, and CNR at the optimal monoenergetic level.
More Related Videos
10:50Osmotic Drug Delivery to Ischemic Hindlimbs and Perfusion of Vasculature with Microfil for Micro-Computed Tomography Imaging
Published on: June 29, 2013
10:16Retrograde Perfusion and Filling of Mouse Coronary Vasculature as Preparation for Micro Computed Tomography Imaging
Published on: February 10, 2012
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Virtual Work
In static equilibrium, a body can experience an imaginary or virtual movement, such as displacement or rotation. The virtual work done by a force is equal to the dot product of force and virtual displacement in the direction of the force. When it comes to virtually rotating a...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II