Sample Pretreatment Method for Determination of Indicator Polychlorinated Biphenyls in Seafood using Ultrasonic
Bin Zheng1, Hongmei Hu2, Xiaoning Zhang3
1Department of Marine Biological Resource Development and Utilization, Zhejiang Marine Development Research Institute, Zhoushan, China.
Abstract:
A simple, rapid and novel method has been developed and validated for the determination of seven indicator polychlorinated biphenyls (PCBs) in seafood samples by gas chromatography coupled to electron capture detector. Freeze-dried samples were done first ultrasonic extraction by n-hexane:methylene chloride:acetone (3:1:1, v/v), and then one-step clean-up (dispersive solid-phase extraction clean-up) or two-step clean-up (concentrated sulfuric acid purification and dispersive solid-phase extraction clean-up) was selected according to the lipid contents of the samples, if the lipid content was no more than 1%, one-step clean-up was used, otherwise, two-step clean-up was chose. The linearity of this method ranged from 1.25 to 100 μg/L, with regression coefficients ranging between 0.9991 and 0.9998. The limits of detection were in low ng/g level, ranging between 0.005 and 0.0076 ng/g (wet weight). The recoveries of spiked seven PCBs with external calibration method at different concentration levels in Pseudosciaena polyactis, Penaeus vannamei and Sinonovacula constricta were in the range of 78-105%, 73-110% and 75-107%, respectively, and with relative standard deviations of 3.3-5.1%, 3.5-6.3% and 3.4-5.1% (n = 5), respectively. The performance of the proposed method was also compared with traditional soxhlet extraction and column chromatography clean-up on the same real seafood samples and comparable efficiencies were obtained. It is concluded that this method can be successfully applied for the determination of PCBs in different seafood samples.
More Related Videos
09:42Fabrication of a Dipole-assisted Solid Phase Extraction Microchip for Trace Metal Analysis in Water Samples
Published on: August 7, 2016
08:41Integration of Miniaturized Solid Phase Extraction and LC-MS/MS Detection of 3-Nitrotyrosine in Human Urine for Clinical Applications
Published on: July 14, 2017
Related Concept Videos
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Extraction: Advanced Methods
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II