Biotribological Testing and Analysis of Articular Cartilage Sliding against Metal for Implants
Published on: May 14, 2020
Tribological evaluation of biomedical polycarbonate urethanes against articular cartilage
Yusuf Kanca1, Piers Milner1, Daniele Dini2
1The Biomechanics Group, Department of Mechanical Engineering, Imperial College London, London SW7 2AZ, UK.
Abstract:
This research investigated the in-vitro wear and friction performance of polycarbonate urethane (PCU) 80A as they interact with articular cartilage, using a customised multidirectional pin-on-plate tester. Condyles were articulated against PCU 80A discs (Bionate® I and Bionate® II) (configuration 1) and the results arising from these tests were compared to those recorded during the sliding of PCU pins against cartilage plates (configuration 2). Configuration 1 produced steadily increasing coefficient of friction (COF) (up to 0.64 ± 0.05) and had the same trend as the cartilage-on-stainless steel articulation (positive control). When synovial fluid rather than bovine calf serum was used as lubricant, average COF significantly decreased from 0.50 ± 0.02-0.38 ± 0.06 for condyle-on-Bionate® I (80AI) and from 0.41 ± 0.02-0.24 ± 0.04 for condyle-on-Bionate® II (80AII) test configurations (p < 0.05). After 15 h testing, the cartilage-on-cartilage articulation (negative control) tests showed no cartilage degeneration. However, different levels of cartilage volume loss were found on the condyles from the positive control (12.5 ± 4.2 mm3) and the PCUs (20.1 ± 3.6 mm3 for 80 AI and 19.0 ± 2.3 mm3 for 80AII) (p > 0.05). A good correlation (R2 =0.84) was found between the levels of average COF and the volume of cartilage lost during testing; increasing wear was found at higher levels of COF. Configuration 2 showed low and constant COF values (0.04 ± 0.01), which were closer to the negative control (0.03 ± 0.01) and significantly lower than configuration 1 (p < 0.05). The investigation showed that PCU is a good candidate for use in hemiarthroplasty components, where only one of the two articulating surfaces is replaced, as long as the synthetic material is implanted in a region where migrating cartilage contact is achieved. Bionate® II showed better tribological performance, which suggests it is more favourable for use in hemiarthroplasty design.
More Related Videos
Related Concept Videos
Growth of Cartilage and Bone Tissue
Self-Evaluation: Self-Enhancement and Self-Verification
Nursing Evaluation
Self-Evaluation Maintenance Model
Dysrhythmias V: Evaluating Dysrhythmias
Social Foundations of Self III: Self-Evaluation