Reduction of seed motion using a bio-absorbable polymer coating during permanent prostate brachytherapy using a mick
Gregory R Warrell1, Yan Xing2, Tarun K Podder1
1Department of Radiation Oncology, University Hospitals Cleveland Medical Center, Cleveland, OH, USA.
The addition of a vicryl coating to radioactive seeds significantly reduces seed movement during low dose rate (LDR) prostate brachytherapy. This innovation may improve treatment consistency and reduce complications associated with seed migration.
Area of Science:
- Oncology
- Medical Physics
- Radiotherapy
Background:
- Low dose rate (LDR) prostate brachytherapy utilizes radioactive seeds for cancer treatment.
- Seed migration can lead to suboptimal radiation delivery and potential complications.
- A bio-absorbable vicryl coating has been developed to enhance seed stability.
Purpose of the Study:
- To evaluate the efficacy of a vicryl coating in preventing seed movement during LDR prostate brachytherapy.
- To compare the incidence and severity of seed slippage between coated and uncoated radioactive seeds.
- To assess the impact of seed coating on initial implant stability.
Main Methods:
- A single-institution study involving 47 patients undergoing LDR prostate brachytherapy.
- Patients received either coated (n=26) or uncoated (n=21) AgX100 125I or uncoated Model 200 103Pd seeds.
- Ultrasound-guided implantation allowed real-time categorization of seed position (in place, moved slightly, or out of view).
Main Results:
- Coated seeds showed a significantly lower incidence of movement (3.1%) compared to uncoated seeds (6.9%), representing a 54.4% reduction (P=0.0026).
- Coated AgX100 seeds had a 66.7% lower movement rate than uncoated AgX100 seeds (P=0.038).
- No significant differences in Day-0 postoperative dosimetric coverage were observed, though further analysis is needed.
Conclusions:
- Vicryl coating significantly reduces the propensity of radioactive seeds to slip from their intended positions.
- This improved stability may help maintain dosimetric integrity in LDR prostate brachytherapy.
- Further investigation into postoperative dosimetry with coated seeds is warranted.
More Related Videos
10:27Preparation of Highly Porous Coordination Polymer Coatings on Macroporous Polymer Monoliths for Enhanced Enrichment of Phosphopeptides
Published on: July 14, 2015
09:28Engineering Molecular Recognition with Bio-mimetic Polymers on Single Walled Carbon Nanotubes
Published on: January 10, 2017
Related Concept Videos
Dynamics Of Circular Motion: Applications
Application of Antiderivatives: Linear Motion
Seed Structure and Early Development of the Sporophyte
Polymers
Polymers
Oxidation-Reduction Reactions