Optical Coherence Tomography Minimum Intensity as an Objective Measure for the Detection of Hydroxychloroquine
Ali M Allahdina1, Paul F Stetson2, Susan Vitale1
1National Eye Institute, National Institutes of Health, Bethesda, Maryland, United States.
Purpose:
As optical coherence tomography (OCT) minimum intensity (MI) analysis provides a quantitative assessment of changes in the outer nuclear layer (ONL), we evaluated the ability of OCT-MI analysis to detect hydroxychloroquine toxicity.
Methods:
Fifty-seven predominantly female participants (91.2% female; mean age, 55.7 ± 10.4 years; mean time on hydroxychloroquine, 15.0 ± 7.5 years) were enrolled in a case-control study and categorized into affected (i.e., with toxicity, n = 19) and unaffected (n = 38) groups using objective multifocal electroretinographic (mfERG) criteria. Spectral-domain OCT scans of the macula were analyzed and OCT-MI values quantitated for each subfield of the Early Treatment Diabetic Retinopathy Study (ETDRS) grid. A two-sample U-test and a cross-validation approach were used to assess the sensitivity and specificity of toxicity detection according to OCT-MI criteria.
Results:
The medians of the OCT-MI values in all nine of the ETDRS subfields were significantly elevated in the affected group relative to the unaffected group (P < 0.005 for all comparisons), with the largest difference found for the inner inferior subfield (P < 0.0001). The receiver operating characteristic analysis of median MI values of the inner inferior subfields showed high sensitivity and high specificity in the detection of toxicity with area under the curve = 0.99.
Conclusions:
Retinal changes secondary to hydroxychloroquine toxicity result in increased OCT reflectivity in the ONL that can be detected and quantitated using OCT-MI analysis. Analysis of OCT-MI values demonstrates high sensitivity and specificity for detecting the presence of hydroxychloroquine toxicity in this cohort and may contribute additionally to current screening practices.