Accuracy evaluation of distance inverse square law in determining virtual electron source location in Siemens Primus
Hamid Shafaei Douk1, Mahmoud Reza Aghamiri1, Mahdi Ghorbani2
1Radiation Medicine Department, Faculty of Shahid Beheshti University, Tehran, Iran.
Aim:
The aim of this study is to evaluate the accuracy of the inverse square law (ISL) method for determining location of virtual electron source (SVir) in Siemens Primus linac.
Background:
So far, different experimental methods have presented for determining virtual and effective electron source location such as Full Width at Half Maximum (FWHM), Multiple Coulomb Scattering (MCS), and Multi Pinhole Camera (MPC) and Inverse Square Law (ISL) methods. Among these methods, Inverse Square Law is the most common used method.
Materials And Methods:
Firstly, Siemens Primus linac was simulated using MCNPX Monte Carlo code. Then, by using dose profiles obtained from the Monte Carlo simulations, the location of SVir was calculated for 5, 7, 8, 10, 12 and 14 MeV electron energies and 10 cm × 10 cm, 15 cm × 15 cm, 20 cm × 20 cm and 25 cm × 25 cm field sizes. Additionally, the location of SVir was obtained by the ISL method for the mentioned electron energies and field sizes. Finally, the values obtained by the ISL method were compared to the values resulted from Monte Carlo simulation.
Results:
The findings indicate that the calculated SVir values depend on beam energy and field size. For a specific energy, with increase of field size, the distance of SVir increases for most cases. Furthermore, for a special applicator, with increase of electron energy, the distance of SVir increases for most cases. The variation of SVir values versus change of field size in a certain energy is more than the variation of SVir values versus change of electron energy in a certain field size.
Conclusion:
According to the results, it is concluded that the ISL method can be considered as a good method for calculation of SVir location in higher electron energies (14 MeV).
Related Concept Videos
Sources of Law
Constitutional law is foundational, deriving from federal and state constitutions, and...
Electronic Distance Measuring Instruments
Chemical Stoichiometry and Gases: Using Ideal Gas Law to Determine Moles
Second Law of Thermodynamics
Distance Problem
Concentration and Rate Law
For example, in a generic reaction aA + bB ⟶ products, where a and b are stoichiometric coefficients, the rate law can be written as: