Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Advanced image processing with fusion and calcification enhancement in transcatheter aortic valve implantation:
Pavel Overtchouk1, Arnaud Sudre1, Cédric Delhaye1
1Department of Cardiology and Cardiovascular Surgery, Institut Coeur Poumon, Centre Hospitalier Regional et Universitaire de Lille, Lille, France.
Objectives:
Radiation exposure is a concern for both patients and operators during transcatheter aortic valve implantation (TAVI). Efforts to reduce radiation dose are warranted. We aimed at investigating if per-operative advanced image processing can reduce patient and operator irradiation use during TAVI.
Methods:
We performed a prospective single-centre observational study comparing patient and operator radiation exposure using standard fluoroscopy (control group) or a novel technology of live advanced fluoroscopic image processing (test group) among consecutive patients undergoing TAVI between August 2015 and April 2016. Patient irradiation (dose-area product, effective dose and air kerma), contrast media volume and clinical outcomes were assessed.
Results:
Among 152 elderly [median age (interquartile range): 83 (78-87)] patients (n = 76 per group) undergoing TAVI, baseline clinical characteristics were similar between the control and test groups, except for a higher median EuroSCORE II (2.8% vs 2.3%, P = 0.02) and higher rate of TAVI for failing surgical bioprosthesis (11.8% vs 2.6%, P = 0.03) in the control group. The dose-area product was reduced in the test group: mean reduction of -27.5 Gy × cm2 [95% confidence intervals (CIs): 15.9-39.1, P < 0.001]. Furthermore, effective dose [mean reduction -6.5 (95% CI: 5.9-7.2) mSv, P < 0.001] and air kerma [mean reduction -167.5 (95% CI 163.4-177.3) mGy, P < 0.001] were lower in the test group. Fluoroscopy time, contrast volume and clinical outcomes were similar.
Conclusions:
Patient radiation exposure was significantly reduced using a novel live advanced fluoroscopy image processing with calcification enhancement and fusion of the virtual aortic annulus without compromising patient safety.
Related Concept Videos
Biological Effects of Radiation
Nuclear Fusion
A helium nucleus has a mass that is 0.7% less than that of four hydrogen nuclei; this lost mass is converted into energy during the fusion. This reaction produces about...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Radiation: Applications
The average...
Impact of Groups on Groups
Absorption of Radiation