Masayuki Umemura1, Kanae Maegawa1, Daichi Arai1
11Laboratory of Pharmacy Practice and Sciences, School of Pharmacy, Aichi Gakuin University, 1-100 Kusumoto-cho, Chikusa-ku, Nagoya, 464-8650 Japan.
Background:
Infusion sets designed for peristaltic finger smart pumps (PFSPs) are necessary for the pumps' accurate handling. We previously found that medication dispensing is occasionally incomplete following the calculated infusion time when using certain combinations of PFSPs and infusion sets at a Japanese hospital. Thus, in this study, we investigated the cause of this observed delay by determining the effect of infusion set attachment technique on dispensing time using a combination of three kinds of PFSPs and five kinds of polyvinyl chloride (PVC) and polybutadiene (PB) infusion sets.
Methods:
PFSPs with their exclusive infusion sets were used. The PVC and PB infusion sets were either not stretched or stretched to 1-3 cm and attached to the PFSP's liquid delivery system. PFSP dispensing rates were set at 25-400 mL/h. The primary outcome was the time required to dispense 100 mL of saline in a volumetric flask.
Results:
The complete dispensing time correlated with the input time for all equipment combinations when the infusion sets were not stretched before attachment to the PFSP (R2 = 0.9998-1.0000). When stretched, the complete dispensing time was longer than the input time (P < 0.01-0.05, analysis of variance with Tukey-Kramer multiple comparisons). The maximum dispensing time extension ratio for the PVC and PB infusion sets was 141.8% and 113.0%, respectively.
Conclusion:
Certain attachment techniques for infusion sets can adversely prolong drug dispensing time. As such, pharmacists should provide medical staff with information about the devices used to administer drugs, as well as about the drugs themselves.
Related Concept Videos
Setting Time of Cement
Attachment
Theory of Romantic Attachment in Adulthood
ATP Driven Pumps III: V-type Pumps
The peripheral or cytosolic V1 domain with eight subunits is involved in ATP hydrolysis. The integral or transmembrane V0 domain containing at least five subunits...
ATP Driven Pumps II: P-type Pumps
A typical P-type pump has three cytosolic domains: nucleotide-binding (N), phosphorylation (P), and activator (A) domains. These domains are connected to the membrane-spanning helices by short amino acid segments. ATP hydrolysis and covalent phosphoenzyme intermediate formation are crucial parts of the catalytic cycle. At the highly...
Attachment Styles