Human Fetal Blood Flow Quantification with Magnetic Resonance Imaging and Motion Compensation
Published on: January 7, 2021
Intra-voxel incoherent motion MRI of the living human foetus: technique and test-retest repeatability
András Jakab1,2, Ruth Tuura1, Raimund Kottke3
11Center for MR-Research, University Children's Hospital, Steinwiesstrasse 75, 8032 Zurich, Switzerland.
Background:
Our purpose was to test the within-subject (test-retest) reproducibility of the perfusion fraction, diffusion coefficient, and pseudo-diffusion coefficient measurements in various foetus organs and in the placenta based on the intra-voxel incoherent motion (IVIM) principle.
Methods:
In utero diffusion-weighted IVIM magnetic resonance imaging (MRI) was performed in 15 pregnant women (pregnancy age 21-36 weeks) on 1.5-T and 3.0-T clinical scanners with b-factors in the range of 0-900 s/mm2 in 16 steps. A bi-exponential model was fitted on the volume-averaged diffusion values. Perfusion fraction (f), diffusion coefficient (d), and pseudo-diffusion coefficient (D*) were calculated. Within-subject reproducibility was evaluated as test-retest variability (VAR %) of the IVIM parameters in the foetal frontal cortex, frontal white matter, cerebellum, lungs, kidneys, liver, and in the placenta.
Results:
For the foetal lungs, liver and the placenta, test-retest variability was in the range of 14-20% for f, 12-14% for d, and 17-25% for D*. The diffusion coefficients of the investigated brain regions were moderately to highly reproducible (VAR 5-15%). However, f and D* showed inferior reproducibility compared to corresponding measures for the lungs, liver, and placenta. The IVIM parameters of the foetal kidney were revealed to be highly variable across scans.
Conclusions:
IVIM MRI potentially provides a novel method for examining microvascular perfusion and diffusion in the developing human foetus. However, reproducibility of perfusion and diffusion parameters depends greatly upon data quality, foetal and maternal movements, and foetal-specific image post-processing.
Related Concept Videos
Development of the Sexual Organs in the Embryo and Fetus
Near the gonadal ridges, two duct systems are present: the mesonephric ducts (Wolffian ducts) and paramesonephric ducts (Müllerian ducts). These ducts form the basis for the...
Equation of Motion: General Plane motion
Moreover, the body's center of mass experiences a rotational effect as a result of these couple moments. This rotation can be articulated as the...
Absolute Motion Analysis- General Plane Motion
As the drone's propellers rotate, an upward force is generated that counteracts the force of gravity, enabling the drone to lift off from the ground. This initial movement of the drone is along a straight path, representing a form of translational motion. In this phase, every point on the...
Equation of Motion: General Plane motion - Problem Solving
The friction between the roller and the ground is characterized by two coefficients. The static friction coefficient is 0.15, while the kinetic friction coefficient is 0.1. These values are crucial in understanding the interaction between...
Projectile Motion: Example
When we speak of the range (R) of a projectile on...
Simple Harmonic Motion and Uniform Circular Motion
There is an easy way to produce simple harmonic motion by using uniform circular motion. For instance, consider a ball attached to a uniformly rotating...