Urine Reagent Strips Are Inaccurate for Assessing Hypohydration: A Brief Report
J D Adams1, Catalina Capitan-Jiménez2, Robert A Huggins3
1Hydration Science Lab, University of Arkansas, Fayetteville, Arkansas.
Objective:
To assess the diagnostic ability of urine reagent strips to identify hypohydration based on urine specific gravity (USG).
Design:
This study examined the agreement of USG between strips and refractometry with Bland-Altman, whereas the diagnostic ability of the strips to assess hypohydration was performed by receiver operating characteristic analysis.
Setting:
Arkansas high school football preseason practice.
Participants:
Four hundred fourteen fresh urine samples were analyzed.
Main Outcome Measures:
Urine specific gravity was assessed by both reagent strips and refractometry. Cutoffs of >1.020 and >1.025 were used for identifying hypohydration.
Results:
Bland-Altman analysis showed agreement of the 2 methods. Overall diagnostic ability of the urine strip to identify hypohydration was fair (area under the curve 72%-78%). However, the sensitivity to correctly identify hypohydration was poor (63%-71%), and the specificity of correctly identifying euhydration was poor to fair (68%-83%).
Conclusion:
The urine strip method is not valid for assessing hypohydration.
Related Concept Videos
Filtration and Urine Formation
Urine Studies II: Urine Culture and Sensitivity Test
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Data Reporting and Recording
EDTA: Auxiliary Complexing Reagents
Urine Studies I: Urinalysis