Related Experiment Video
Updated: Aug 1, 2026

Label-free Single Molecule Detection Using Microtoroid Optical Resonators
Published on: December 29, 2015
Metrological 2iOF fibre-optic system for position and displacement measurement with 31 pm resolution
Karolina Orłowska1, Michał Świątkowski1, Piotr Kunicki1
1Faculty of Microsystem Electronics and Photonics, Nanometrology Division, Wroclaw University of Science and Technology, Janiszewskiego 11/17, 50-372 Wroclaw, Poland.
Abstract:
In the present paper, we describe a high sensitivity intensity fibre-optic displacement sensor with tens of picometre resolution combined with a sub-picometre resolution interferometric calibration system. Both integrated components form the so-called "2 in one ferrule" system 2iOF. The design and construction of the presented device depend on integrating two sensors' systems within one fibre-optic measuring head, which allows performing in situ calibration process with no additional time-consuming adjustment procedure. The resolution of the 2iOF system is 31 pm/Hz1/2 obtained with an interferometric Fabry-Perot based calibration system-providing accuracy better than tens of fm/Hz1/2 within 1 MHz bandwidth in the measurement range of up to 100 μm. The direct response from the intensity sensor is then the 2iOF output one. It is faster and more convenient to analyze in comparison, with much better resolution (3 orders of magnitude higher) but on the other hand also more time consuming and dependent on the absolute sample position interferometer. The proposed system is flexible and open to various applications. We will present the results of the piezoelectrical actuator displacement measurements, which were performed using the developed system.