Biomechanics of two external fixator devices used in rat femoral fractures
Liza Osagie-Clouard1, Joshua Kaufmann1, Gordon Blunn1,2
1Division of Surgery, University College London, Stanmore, UK.
Abstract:
The use of external fixators allows for the direct investigation of newly formed interfragmentary bone, and the radiographic evaluation of the fracture. We validated the results of a finite element (FE) model with the in vitro stiffness' of two widely used external fixator devices used for in vivo analysis of fracture healing in rat femoral fractures with differing construction (Ti alloy ExFix1 and PEEK ExFix2). Rat femoral fracture fixation was modeled using two external fixators. For both constructs an osteotomy of 2.75 mm was used, and offset maintained at 5 mm. Tufnol, served as standardized substitutes for rat femora. Constructs were loaded under axial compression and torsion. Overall axial and torsional stiffness were compared between the in vitro models and FE results. FE models were also used to compare the fracture movement and overall pattern of von Mises stress across the external fixators. In vitro axial stiffness of ExFix1 was 29.26 N/mm ± 3.83 compared to ExFix2 6.31 N/mm ± 0.67 (p* < 0.05). Torsional stiffness of ExFix1 was 47.5 Nmm/° ± 2.71 compared to ExFix2 at 19.1 Nmm/° ±1.18 (p* < 0.05). FE results predicted similar comparative ratios between the ExFix1 and 2 as the in vitro studies. FE results predicted considerably larger interfragmentary motion in the ExFix2 comparing to ExFix1. We demonstrated significant differences in the stiffness' of the two external fixators as one would expect from such variable designs; yet, importantly we validated the utility of an FE model for the analysis and prediction of changes in fracture mechanics dependent on fixator choice. © 2018 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 37:293-298, 2019.
Related Concept Videos
Work Done on a System by External Force
In the presence of a non-conservative opposing force, like friction, some part of the work done...
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Internal and External Forces
External and Internal Respiration
General External Flow Characteristics
External Anatomy of the Kidney
The kidneys are located in the retroperitoneal space on either side of the vertebral column, protected posteriorly by the 11th and 12th ribs. The right kidney sits slightly lower than the left owing to the presence of the liver...