In vitro comparison of the adsorption of inflammatory mediators by blood purification devices
Benjamin Malard1, Corine Lambert2, John A Kellum3
1R&D Department, Gambro Industries, 7 avenue Lionel Terray, 69330, Meyzieu, France. benjamin_malard@baxter.com.
Background:
Septic shock, a leading cause of acute kidney injury, induces release of pro-/anti-inflammatory mediators, leading to increased mortality and poor renal recovery. This is the first in vitro study directly comparing three single-use blood purification devices in terms of removing sepsis-associated mediators and endotoxins.
Methods:
In vitro hemoperfusion was performed using oXiris®, CytoSorb®, and Toraymyxin®. Heparinized human plasma from healthy volunteers was pre-incubated with pathologic quantities of inflammatory mediators and filtered in a closed-loop circulation model for 2 h. For each device, the removal of 27 inflammatory mediators was measured over time. Endotoxin removal mediated by oXiris and Toraymyxin was assessed using hemoperfusion over 6 h.
Results:
Endotoxin (lipopolysaccharide) removal was most rapid with Toraymyxin; mean adsorptive clearance over the first 30 min was ~ 20 ml/min vs ~ 8 ml/min with oXiris (p < 0.05). There was minimal endotoxin removal with CytoSorb (1 ml/min). At 120 min, there was no significant difference between the endotoxin removal rates using oXiris (mean ± standard deviation, 68.0 ± 4.4%) and Toraymyxin (83.4 ± 3.8%); both were significantly higher vs CytoSorb (- 6.3 ± 4.9%; p < 0.05). Total removal with oXiris was 6.9 μg vs 9.7 μg for Toraymyxin, where the total lipopolysaccharide quantity introduced was approximately 15.8 μg. Removal rates of pro-/anti-inflammatory cytokines and other inflammatory mediators were similar between oXiris and CytoSorb and were higher with CytoSorb and oXiris vs Toraymyxin. Granulocyte colony-stimulating factor was only effectively adsorbed by CytoSorb (99.4%). Differences were detected between the adsorption mechanism of the devices; binding to oXiris was mainly ionic, while CytoSorb was hydrophobic. No specific protein adsorption was found qualitatively with Toraymyxin.
Conclusions:
Adsorption rate kinetics varied for individual inflammatory mediators using the three blood purification devices. Mechanisms of adsorption differed between the devices. oXiris was the only device tested that showed both endotoxin and cytokine removal. oXiris showed similar endotoxin adsorption to Toraymyxin and similar adsorption to CytoSorb for the removal of other inflammatory mediators. Differences in device removal capacities could enable treatment to be more tailored to patients.
Related Concept Videos
Analyte Adsorption and Distribution
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Inflammatory Response
Inflammation can be triggered by various stimuli, such as impact, abrasion, chemical irritation, infections, and extreme hot or cold temperatures. These can damage cells and connective tissue fibers,...
Inflammatory Response II: Inflammatory Exudate and Tissue Repair
The typical wound exudate is odorless, transparent, straw-colored, thin, and watery. Exudate, however, can differ depending on the state of wound healing. Likewise, the...
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons