Comparing Change in Anterior Curvature After Corneal Cross-linking Using Scanning-slit and Scheimpflug Technology
Paul Z Lang1, Praneetha Thulasi2, Sumitra S Khandelwal3
1USC Roski Eye Institute, Keck School of Medicine of the University of Southern California, Los Angeles, California, USA.
Purpose:
To evaluate the correlation between anterior axial curvature difference maps following corneal cross-linking (CXL) for progressive keratoconus obtained from Scheimpflug-based tomography and Placido-based topography.
Design:
Between-devices reliability analysis of randomized clinical trial data.
Methods:
Corneal imaging was collected at a single-center institution preoperatively and at 3, 6, and 12 months postoperatively using Scheimpflug-based tomography (Pentacam; Oculus Inc, Lynnwood, Washington, USA) and scanning-slit, Placido-based topography (Orbscan II; Bausch & Lomb, Rochester, New York, USA) in patients with progressive keratoconus receiving standard protocol CXL (3 mW/cm2 for 30 minutes). Regularization index (RI), absolute maximum keratometry (K Max), and change in K Max (ΔK Max) were compared between the 2 devices at each time point.
Results:
Fifty-one eyes from 36 patients were evaluated at all time points. Values were significantly different at all time points (56.01 ± 5.3 diopters [D] Scheimpflug vs 55.04 ± 5.1 D scanning-slit preoperatively [P = .003]; 54.58 ± 5.3 D Scheimpflug vs 53.12 ± 4.9 D scanning-slit at 12 months [P < .0001]) but strongly correlated between devices (r = 0.90-0.93) at all time points. The devices were not significantly different at any time point for either ΔK Max or RI but were poorly correlated at all time points (r = 0.41-0.53 for ΔK Max, r = 0.29-0.48 for RI). At 12 months, 95% limits of agreement were 7.51 D for absolute K Max, 8.61 D for ΔK Max, and 19.86 D for RI.
Conclusions:
Measurements using Scheimpflug and scanning-slit Placido-based technology are correlated but not interchangeable. Both devices appear reasonable for separately monitoring the cornea's response to CXL; however, caution should be used when comparing results obtained with one measuring technology to the other.
More Related Videos
04:48Establishing a Severe Corneal Inflammation Model in Rats Based on Corneal Epithelium Curettage Combined with Corneal Sutures
Published on: November 22, 2024
07:29Three Different Protocols of Corneal Collagen Crosslinking in Keratoconus: Conventional, Accelerated and Iontophoresis
Published on: November 12, 2015
Related Concept Videos
Degree of Curvature and Radius of Curvature
Crossing Over
The homologous pairs of sister chromosomes—one from the maternal and one from the paternal genome—then begin to align alongside each other lengthwise, matching corresponding DNA positions in a process...
Cross-Sectional Research
Vertebral Column: Regions and Curvature
Regions of the Vertebral Column
In an adult, the spine is subdivided into five regions: the cervical, the thoracic, the lumbar, the sacral, and the coccygeal region. The spine initially develops as a series of 33 vertebrae; after 20 years of age, the nine bones in the sacral region, five sacral, and four coccygeal bones fuse to form...
Covalently Linked Protein Regulators
These groups modify specific amino acids in a protein....
Leaky Scanning