Influence of k-space trajectory corrections on proton density mapping with ultrashort echo time imaging: Application
Peter Latta1, Zenon Starčuk2, Marco L H Gruwel3
1Central European Institute of Technology, Masaryk University, Brno, Czech Republic.
Purpose:
To evaluate the impact of MR gradient system imperfections and limitations for the quantitative mapping of short T2* signals performed by ultrashort echo time (UTE) acquisition approach.
Materials And Methods:
The measurement of short T2* signals from a phantom and a healthy volunteer study (8 subjects of average age 28 ± 4 years) were performed on a 3T scanner. The characteristics of the gradient system were obtained using calibration method performed directly on the measured subject or phantom. This information was used to calculate the actual sampling trajectory with the help of a parametric eddy current model. The actual sample positions were used to reconstruct corrected images and compared with uncorrected data.
Results:
Comparison of both approaches, i.e., without and with correction of k-space sampling trajectories revealed substantial improvement when correction was applied. The phantom experiments demonstrate substantial in-plane signal intensity variations for uncorrected sampling trajectories. In the case of the volunteer study, this led to significant differences in relative proton density (RPD) estimation between the uncorrected and corrected data (P = 0.0117 by Wilcoxon matched-pairs test) and provides for about ~15% higher values for short T2* components of white matter (WM) in the case of uncorrected images.
Conclusion:
The imperfection of the applied gradients could induce errors in k-space data sampling which further propagates into the fidelity of the UTE images and jeopardizes precision of quantification. However, the study proved that measurement of gradient errors together with correction of sample positions can contribute to increased accuracy and unbiased characterization of short T2* signals.
Related Concept Videos
Echo
Imagine the sound is reflected back to the ears. Assuming that the source is very close to the human, the difference between hearing the two sounds—the emitted sound and the reflected sound—may be more than the minimum time for perceiving distinct sounds. If this is the case,...
Classifying Matter by State
Classifying Matter by Composition
According to its composition, the matter can be classified into two broad categories — pure substances and mixtures.
A pure substance is a form of matter that has a constant composition throughout with uniform properties. For example, any sample of sucrose has the same composition and same physical properties, such as melting point, color, and sweetness, regardless of the source from which it is isolated.
A mixture is composed of two or...
Space-Time Curvature and the General Theory of Relativity
This has been verified in many experiments. However, space and time are no longer absolute. Two observers moving relative to one another do not agree on the length of objects or the passage of time. The mechanics of objects based on Newton's laws of...
Orthogonal Trajectories
Physical and Chemical Properties of Matter