Cadaveric Study of the Safety and Device Functionality of Magnetically Controlled Growing Rods After Exposure to
Selina Poon1, Yen Hsun Chen2, Stephen F Wendolowski2
1Department of Orthopaedics, Shriners for Children Medical Center, 909 South Fair Oaks Avenue, Pasadena, CA 91105, USA.
Study Design:
Cadaveric study.
Objective:
To establish the safety and efficacy of magnetically controlled growing rods (MCGRs) after magnetic resonance imaging (MRI) exposure.
Summary Of Background Data:
MCGRs are new and promising devices for the treatment of early-onset scoliosis (EOS). A significant percentage of EOS patients have concurrent spinal abnormalities that need to be monitored with MRI. There are major concerns of the MRI compatibility of MCGRs because of the reliance of the lengthening mechanism on strongly ferromagnetic actuators.
Methods:
Six fresh-frozen adult cadaveric torsos were used. After thawing, MRI was performed four times each: baseline, after implantation of T2-T3 thoracic rib hooks and L5-S1 pedicle screws, and twice after MCGR implantation. Dual MCGRs were implanted in varying configurations and connected at each end with cross connectors, creating a closed circuit to maximize MRI-induced heating. Temperature measurements and tissue biopsies were obtained to evaluate thermal injury. MCGRs were tested for changes to structural integrity and functionality. MRI images obtained before and after MCGR implantation were evaluated.
Results:
Average temperatures increased incrementally by 1.1°C, 1.3°C, and 0.5°C after each subsequent scan, consistent with control site temperature increases of 1.1°C, 0.8°C, and 0.4°C. Greatest cumulative temperature change of +3.6°C was observed adjacent to the right-sided actuator, which is below the 6°C threshold cited in literature for clinically detectable thermal injury. Histologic analysis revealed no signs of heat-induced injury. All MCGR actuators continued to function properly according to the manufacturer's specifications and maintained structural integrity. Significant imaging artifacts were observed, with the greatest amount when dual MCGRs were implanted in standard/offset configuration.
Conclusions:
We demonstrate minimal MRI-induced temperature change, no observable thermal tissue injury, preservation of MCGR-lengthening functionality, and no structural damage to MCGRs after multiple MRI scans. Expectedly, the ferromagnetic actuators produced substantial MR imaging artifacts.
Level Of Evidence:
Level V.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies IV: Magnetic Resonance Imaging
Atomic Nuclei: Magnetic Resonance
Nuclear Magnetic Resonance (NMR): Overview
NMR spectroscopy generates a spectrum where the characteristic absorption frequencies of the sample are...
Colors and Magnetism
When atoms or molecules absorb light at the proper frequency, their electrons are excited to higher-energy orbitals. For many main group atoms and molecules, the absorbed photons are in the ultraviolet range of the electromagnetic spectrum, which cannot be detected by the human eye. For coordination compounds, the energy difference between the d orbitals often allows photons in the visible range to be absorbed and emitted, which is seen as colors by the human...
Magnetism
An individual magnetic pole cannot be isolated. No matter how small, every piece of a magnet contains a north pole and a south...