Related Experiment Video
Updated: Feb 10, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Inadvertent defibrillator lead placement into the left ventricle after MitraClip implantation: A case report
Giuseppe Santarpia1, Francesco Passafaro, Eugenia Pasceri
1Division of Cardiology, Department of Medical and Surgical Sciences, "Magna Graecia" University URT-CNR, Department of Medicine, Consiglio Nazionale delle Ricerche, Catanzaro, Italy.
Rationale:
Inadvertent pacemaker/defibrillator lead placement into the left ventricle is an unusual cardiac device-related complication and its diagnosis is not always easy and often misunderstood. Thromboembolic events are frequently associated with this procedural complication. Percutaneous lead extraction should be performed when diagnosis is made early after device implantation while long-life oral anticoagulation is a wise option when the diagnosis is delayed and the lead is not removed.
Patient Concerns:
A 65-year-old man affected by dilated cardiomyopathy, previously treated with a percutaneous mitral valve repair, with 2 MitraClip devices, and later with dual chamber cardioverter/defibrillator implantation, returned in outpatient clinics 2 months after discharge for deterioration of dyspnea; transthoracic echocardiography revealed that the shock lead had been accidentally placed in the apex of the left ventricle.
Diagnoses:
The unintentional lead malposition through the iatrogenic atrial septal defect and its presence into the mitral valve orifice, together with the 2 clip devices implanted, generated an acceleration of transvalvular diastolic flow, determining a moderate stenosis of the mitral valve, as well as promoting a worsening of the degree of valvular regurgitation.
Interventions:
Oral anticoagulation therapy was started and a mechanical lead extraction was percutaneously performed. A new defibrillator lead was later appropriately positioned in the apex of the right ventricle.
Outcomes:
The patient was discharged 3 days after intervention and the follow-up, performed 1 month after discharge, was uneventful.
Lessons:
Complex interventional procedures and implantation of multiple devices can increase procedural troubles and the risk of mechanical complications related to pacemaker/defibrillator implantation. Careful observation of the QRS complex morphology on the electrocardiogram (ECG), during paced rhythm, and the achievement of the echocardiographic examination, in the postprocedural phase, allow an early diagnosis of lead malposition.
Related Concept Videos
Anatomy of the Brain: Ventricles
Data Reporting and Recording
Phase-lead and Phase-lag Controllers
Types of Reports I: Hands-off Report
Following are the key components and categories of hand-off reports:
Purpose and Process:
Types of Reports II: Incident or Occurrence Report
Purposes:
In the healthcare industry, reports play a crucial role in documenting incidents within an agency. The primary objective of these reports is to ensure patient safety, uphold the...
Types of Reports III: Telephone and Verbal Reports
Here's an overview of each type:
Telephone Orders