Related Experiment Video
Updated: Feb 10, 2026

Imaging of the Microstructural Failure Mechanism in the Human Hip
Published on: September 29, 2023
Mechanical failure of a femoral lengthening nail: a case report
134800 Bob Wilson Drive, San Diego, CA 92134 USA.
Background:
This case report describes the mechanical failure of a femoral lengthening nail in order to allow early recognition and prevent its occurrence.
Case Presentation:
A femoral lengthening nail was used to achieve goal distraction length over eight weeks and sustained a mechanical failure during progressive weightbearing. Re-distraction was attempted with subsequent non-weight-bearing status. Although the patient was not able to attain the original goal length, radiographs demonstrated stable shortening of the patient's right lower extremity along with early signs of incomplete regenerate bone. She eventually received exchange nailing with bone graft.
Conclusion:
Post-operative protocol and weight-bearing status should be further researched and standardized after re-evaluating the mechanical stiffness of the intramedullary nail construct.
Related Concept Videos
Accessory Structures of the Skin: Nails
The main components of a nail include the following.
Nail Plate: The nail plate is the visible portion of the nail that extends beyond the fingertips or toes. It is a hard, translucent...
Data Reporting and Recording
Types of Reports I: Hands-off Report
Following are the key components and categories of hand-off reports:
Purpose and Process:
Types of Reports II: Incident or Occurrence Report
Purposes:
In the healthcare industry, reports play a crucial role in documenting incidents within an agency. The primary objective of these reports is to ensure patient safety, uphold the...
Pathophysiology of Heart Failure
Heart Failure I: Introduction