Accuracy of oscillatory pressure measured by mechanical ventilators during high frequency oscillatory ventilation in
Chiara Veneroni1, Marta Tossici1, Emanuela Zannin1
1Dipartimento di Elettronica, Informazione e Bioingegneria, Politecnico di Milano University, Milano, Italy.
Background:
Oscillatory pressure (ΔP) measurement during high frequency oscillatory ventilation (HFOV) is technically challenging and influenced by all the components of the measurement system.
Objectives:
To evaluate the differences between the ΔP delivered at the inlet of the endotracheal tube and those displayed by commercial neonatal mechanical ventilators and monitoring devices and to characterize how the ventilator circuit and the flowmeter proximal to the patient affect these differences.
Methods:
Six devices were evaluated while ventilating three mechanical analogues representing the newborn respiratory system in different disease states. ΔP measured at different points of the ventilator circuit were compared.
Results:
ΔP accuracy is highly variable and decreases with increasing oscillation frequency and amplitude, independently of the mean airway pressure. At 15 Hz, a ΔP displayed by ventilators of 40 cmH2 O resulted in a ΔP effectively delivered at the tracheal tube ranging from 22 to 49 cmH2 O, depending on the ventilator model, the ventilator circuit, and the patient condition. At these settings, the errors exclusively due to the ventilator circuit and the presence of the flowmeter ranged from 6 to 9 cmH2 O and from 1 to 6 cmH2 O, respectively.
Conclusions:
The ventilator model, the breathing circuit, the flowmeter, and the patient condition severely impacts ΔP measurement accuracy during HFOV, leading to highly variable performances. This prevents the possibility of using the ΔP required to normalize gas exchange as an indicator of patients' condition, complicates comparison of ventilators performances, and adds a significant element of complexity in clinical management of HFOV.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation I: Indication and Settings
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: