Accelerometer-based atrioventricular synchronous pacing with a ventricular leadless pacemaker: Results from the Micra
Larry Chinitz1, Philippe Ritter2, Surinder Kaur Khelae3
1NYU Langone Medical Center, New York, New York.
Background:
Micra is a leadless pacemaker that is implanted in the right ventricle and provides rate response via a 3-axis accelerometer (ACC). Custom software was developed to detect atrial contraction using the ACC enabling atrioventricular (AV) synchronous pacing.
Objective:
The purpose of this study was to sense atrial contractions from the Micra ACC signal and provide AV synchronous pacing.
Methods:
The Micra Accelerometer Sensor Sub-Study (MASS) and MASS2 early feasibility studies showed intracardiac accelerations related to atrial contraction can be measured via ACC in the Micra leadless pacemaker. The Micra Atrial TRacking Using A Ventricular AccELerometer (MARVEL) study was a prospective multicenter study designed to characterize the closed-loop performance of an AV synchronous algorithm downloaded into previously implanted Micra devices. Atrioventricular synchrony (AVS) was measured during 30 minutes of rest and during VVI pacing. AVS was defined as a P wave visible on surface ECG followed by a ventricular event <300 ms.
Results:
A total of 64 patients completed the MARVEL study procedure at 12 centers in 9 countries. Patients were implanted with a Micra for a median of 6.0 months (range 0-41.4). High-degree AV block was present in 33 patients, whereas 31 had predominantly intrinsic conduction during the study. Average AVS during AV algorithm pacing was 87.0% (95% confidence interval 81.8%-90.9%), 80.0% in high-degree block patients and 94.4% in patients with intrinsic conduction. AVS was significantly greater (P <.001) during AV algorithm pacing compared to VVI in high-degree block patients, whereas AVS was maintained in patients with intrinsic conduction.
Conclusion:
Accelerometer-based atrial sensing is feasible and significantly improves AVS in patients with AV block and a single-chamber leadless pacemaker implanted in the right ventricle.
Related Concept Videos
Simplified Synchronous Machine Model
In this model, each generator is connected to a...
Three-Phase Short Circuit—Unloaded Synchronous Machine
This behavior occurs due to the magnetic flux produced by the short-circuit armature currents. Initially, these currents follow high-reluctance paths but eventually shift to...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Case Studies
Lewis Acids and Bases
A coordinate covalent bond (or dative bond) occurs when one of the atoms in the bond provides both bonding electrons. For example, a coordinate covalent bond occurs when a water molecule combines with a hydrogen ion to form a hydronium ion. A coordinate covalent bond also results when...
Weak Base Solutions