Catheter Ablation in Combination With Left Atrial Appendage Closure for Atrial Fibrillation
Published on: February 26, 2013
Ablation Index and Surround Flow Catheter Irrigation: Impedance-Based Appraisal in Clinical Ablation
Waqas Ullah1, Ross J Hunter2, Malcolm C Finlay2
1Cardiovascular Biomedical Research Unit, Barts Heart Centre, St. Bartholomew's Hospital, Barts Health National Health Service Trust, London, United Kingdom; Cardiology Department, University Hospital Southampton National Health Service Foundation Trust, Southampton, United Kingdom.
Objectives:
This study sought to assess the impact of ablation power and catheter irrigation during clinical radiofrequency ablation using impedance drop.
Background:
In preclinical studies, ablation power and catheter irrigation are determinants of ablation efficacy.
Methods:
Static 30-s left atrial ablations were delivered in patients undergoing their first atrial fibrillation ablation. Impedance drop during ablation (as a measure of efficacy) was compared using the following: the force time integral (FTI); the FTI-P (a cumulative multiple FTI and ablation power), and ablation index (AI), a weighted algorithm including contact force, power, and duration. Comparison was also made between a conventionally irrigated (SmartTouch [ST]) versus surround flow (STSF) contact force-sensing catheter.
Results:
We analyzed 1,013 ablations. For both catheters, the Spearman correlation was higher between impedance drop and AI (rho = 0.89 ST, 0.84 STSF) than FTI-P (rho = 0.71 ST, 0.53 STSF) or FTI (rho = 0.77 ST, 0.52 STSF); p < 0.0005 for each. STSF ablations had lower minimum catheter tip temperatures (25°C [interquartile range (IQR): 25°C to 27°C] vs. 35°C [IQR: 34°C to 36°C]; p < 0.005), and lesser impedance drop per FTI or AI (p < 0.005 for both). For STSF, impedance drop plateaued sooner than for ST with respect to FTI (184g.s vs. 463g.s) and AI (370 AI vs. 430 AI).
Conclusions:
AI is a more complete ablation descriptor than is FTI or FTI-P, reflected by a stronger correlation with impedance drop. STSF ablations have lower impedance drop per AI or FTI than ST ablations do, suggesting different targets should be used if ablating guided by impedance drop with STSF. With ST, ablation beyond 430 AI provides minimal additional biophysical efficacy, suggesting an upper limit to use for clinical ablation.
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Design Example: Design of an Irrigation Channel
Lazarus's Cognitive Appraisal Theory
Primary Appraisal:...
Impedance Combination
Impedances and Admittance
Impedance is a measure of resistance to sinusoidal current flow in an AC circuit. Unlike their behavior in DC circuits, where inductors appear as short circuits and capacitors as open circuits, the behavior of these components in AC...
Series Impedances: Three-Phase Line
Using Kirchhoff's laws, an integro-differential equation for the network is derived. This equation accounts for unbalanced phase currents, which may induce return currents through neutral wires and the earth, seeking the least impedance path. Earth return conductors can replace the...