Benefit of Implantable Cardioverter-Defibrillator Generator Replacement in a Primary Prevention Population-Based
Willy Weng1, John Sapp2, Steve Doucette3
1Department of Medicine, Dalhousie University, Halifax, Nova Scotia, Canada.
Objectives:
This study investigated the benefit of an implantable cardioverter-defibrillator (ICD) generator replacement in patients who did not have an ongoing theoretical indication for ICD therapy at time of replacement.
Background:
Primary prevention ICD therapy is known to reduce mortality in patients with cardiomyopathy and reduced left ventricular systolic function. The data describing outcomes after generator replacement are limited.
Methods:
This was a retrospective cohort study following patients implanted with primary prevention ICD therapy from 2002 until 2015 who subsequently received a generator replacement. Patients with an ongoing theoretical indication for ICD therapy were defined as either left ventricular ejection fraction ≤35% or having had prior appropriate ICD therapy. Outcomes were mortality, appropriate ICD therapy and shock, inappropriate shock, and device and lead complications.
Results:
A total of 614 patients were identified; 173 (28.2%) underwent a generator replacement and were followed for a mean of 2.9 years after replacement; 144 (83.2%) had an ongoing theoretical indication. Patients with no ongoing theoretical indication (n = 29, 16.7%) had lower mortality (hazard ratio [HR]: 0.39, 95% confidence interval [CI]: 0.15-1.00; p = 0.0495), appropriate shock rate (HR: 0.29, 95% CI: 0.09 to 0.96; p = 0.04), and appropriate ICD therapy rate (HR: 0.30, 95% CI: 0.12 to 0.77; p = 0.012) when compared with patients with ongoing theoretical indication. In the entire cohort, there were low rates of inappropriate shock (4.0%), device complication (5.1%), and lead complication (2.3%).
Conclusions:
In patients with primary prevention ICD therapy who underwent generator replacement, improved left ventricular ejection fraction and lack of prior appropriate ICD therapy at time of replacement were associated with a lower risk of mortality and incident ventricular arrhythmia.