Adjustable-Loop Femoral Cortical Suspensory Fixation for Patellar Tendon Anterior Cruciate Ligament Reconstruction: A Time Zero Biomechanical Comparison With Interference Screw Fixation
Related Experiment Video
Updated: Feb 10, 2026

Author Spotlight: Implementing the Enhanced Recovery After Surgery Concept in Rehabilitation Following Anterior Cruciate Ligament Reconstruction
Published on: March 1, 2024
Dayne T Mickelson1, Thomas Lefebvre1, Ken Gall2
1Duke Sports Science Institute, Department of Orthopaedic Surgery, Duke University School of Medicine, Durham, North Carolina, USA.
Background:
Adjustable-loop cortical buttons for femoral fixation of bone-tendon-bone grafts have potential advantages over interference screw fixation; however, these devices have not been benchmarked biomechanically against interference screws. Purpose/Hypothesis: The purpose was to compare the time zero biomechanical properties of commercially available, adjustable-loop cortical button and metallic interference screws for femoral fixation of bone-tendon-bone grafts. It was hypothesized that no significant differences would be found in biomechanical properties between fixation techniques.
Study Design:
Controlled laboratory study.
Methods:
Adjustable-loop cortical buttons (n = 8) and metallic interference screws (n = 8) were used to fix matched pairs of human bone-tendon-bone allografts in porcine distal femurs. These constructs were preconditioned (10 N to 50 N at 1 Hz, 10 cycles), subjected to cyclic loading (50 N to 250 N at 1 Hz, 500 cycles), and then pulled to failure at 20 mm/min.
Results:
The loads to failure (mean ± SD, 700 ± 256 N vs 688 ± 215 N, P = .92) and linear stiffnesses (219 ± 48 N/mm vs 218 ± 49 N/mm, P = .97) for the adjustable-loop cortical button and metallic interference screws, respectively, were not significantly different. Cyclic displacement was higher in the adjustable-loop cortical button group (2.1 ± 0.6 mm vs 1.3 ± 0.4 mm, P = .01). The mechanism of failure was different between groups, with bone block slippage occurring most commonly in the interference screw group (n = 5) and fracture of the bone block through the suture hole occurring most commonly in the adjustable-loop cortical button group (n = 6).
Conclusion:
Adjustable-loop cortical buttons and interference screws have similar time zero failure loads, although cyclic displacement was higher with the adjustable-loop cortical buttons. The mean difference in displacement was less than 1 mm compared with the interference screw.
Clinical Relevance:
Adjustable-loop cortical buttons may be an acceptable alternative to an interference screw for femoral fixation of bone-tendon-bone grafts in anterior cruciate ligament reconstruction. The clinical relevance of the observed differences in cyclic displacement is unknown and should be evaluated in future studies.
More Related Videos
10:32Athymic Rat Model for Evaluation of Engineered Anterior Cruciate Ligament Grafts
Published on: March 26, 2015
06:28Anterior Cruciate Ligament Transection and Synovial Fluid Lavage in a Rodent Model to Study Joint Inflammation and Posttraumatic Osteoarthritis
Published on: September 2, 2025
Related Concept Videos
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...
Carbon-dioxide Fixation
Interference and Diffraction
RNA Interference
This process occurs naturally in cells, often through the activity of genomically-encoded microRNAs. Researchers can take advantage of this mechanism by introducing synthetic RNAs to deactivate specific genes for research or therapeutic purposes. For example, RNAi could be used...
Screw: Problem Solving
To evaluate the...
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.