[Evaluation of the EDOSE for IMRT Delivery Quality Assurance]
Objective:
Evaluation of intensity-modulated radiation therapy(IMRT) delivery quality assurance(DQA) using EDOSE software based on accelerator electronic portal imaging device (EPID).
Methods:
Analysed the dose response uniformity correction of EPID imaging plate in different fields and the gamma pass rates about the fields; And compared the calculated parameters of the EDOSE and the Pinnacle planning station.
Results:
The uniformities on the homogeneous region of thefi elds are good after who corrected by EDOSE. The gamma(γ, 3mm/3%) pass rates are in the ideal range. The volume of the organs at risk(OAR) have no statistical signifi cance except the right lung; A part of the parameters have statistical signifi cance.
Conclusions:
The results of the calculation and analysis of EDOSE has a high agreement with the planning system.