Ultrasound Cyclo-Plasty in Patients with Glaucoma: 1-Year Results from a Multicentre Prospective Study
Giuseppe Giannaccare1, Aldo Vagge2, Stefano Sebastiani3
1Ophthalmology Unit, DIMES, S. Orsola-Malpighi University Hospital, University of Bologna, Bologna, Italy, giuseppe.giannaccare@gmail.com.
Purpose:
The aim of this study was to evaluate the safety and efficacy of ultrasound cyclo-plasty (UCP) for reducing intraocular pressure (IOP) in patients with glaucoma.
Methods:
This is a multicentre prospective study conducted in 3 Italian glaucoma centres. UCP was performed by EyeOP1, which delivers ultrasound beams using 6 piezoelectric transducers activated for 4/6 s (first generation) or 8 s (second generation). Primary outcomes were the mean IOP reduction and the rates of success after 1 year. Secondary outcomes were the mean IOP reduction at each follow-up, and the reduction of the number of hypotensive medications.
Results:
In total, 49 eyes from 47 patients were treated. One year postoperatively, the mean IOP had decreased from 27.7 ± 9.2 to 19.8 ± 6.9 mm Hg (p < 0.001), and the mean number of hypotensive drops and tablets had decreased from 3.2 and 0.5 to 2.3 and 0.2, respectively (p < 0.05). Postoperative IOP reduction was significantly related to preoperative IOP (r2 = 0.5034; p < 0.0001). Second-generation probes determined a significantly higher IOP reduction (p < 0.05). Qualified success was achieved in 25 eyes (51.1%) and complete success in 21 (42.9%), while failure was recorded in 12 (24.5%).
Conclusions:
UCP is safe and effective for reducing IOP. The procedure determines a greater IOP reduction in patients with higher preoperative IOP. Second-generation probes improve outcomes without detrimental effects on safety.
Related Concept Videos
Glaucoma: Overview
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...
Angle Closure Glaucoma: Treatment
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Bioavailability Study Design: Healthy Subjects Versus Patients
SN1 Reaction: Stereochemistry
In the first step of an SN1 reaction, the bond between the electrophilic carbon and the leaving group ionizes to generate the carbocation intermediate. The second step of the mechanism is the nucleophilic attack.
In the formed carbocation, the positively charged carbon is sp2 hybridized with a trigonal planar geometry. As all the three substituents lie on the same plane, a plane of symmetry for the...