Edwards Sapien XT® pulmonic valve compression after resuscitation and successful redilatation
Anja Lehner1, Marinos Kantzis2, Nikolaus A Haas1
1Department for Pediatric Cardiology and Pediatric Intensive Care, LMU - Campus Großhadern, Ludwig Maximilians University, Munich, Germany.
Abstract:
Despite the increasing use of percutaenous valves, little is known about valve performance after external chest compression. We report a case of a compressed pulmonary Edwards Sapien® XT valve after resuscitation. With the patient on ECMO, successful redilatation was performed with unimpaired postprocedural valve function. We aim to increase the awareness for valve distortion after external chest compression with increasing numbers of transcatheter valvulations. We suggest that immediate re-evaluation of the implanted device via biplane fluoroscopy is mandatory after resuscitation in such cases.
More Related Videos
Related Concept Videos
Ecological Succession
Cardiopulmonary Resuscitation I: Adult
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...