Biomechanical Impact of Wrong Positioning of a Dedicated Stent for Coronary Bifurcations: A Virtual Bench Testing
Claudio Chiastra1, Maik J Grundeken2, Carlos Collet2,3
1Laboratory of Biological Structure Mechanics (LaBS), Department of Chemistry, Materials and Chemical Engineering "Giulio Natta", Politecnico di Milano, Piazza Leonardo da Vinci 32, 20133, Milan, Italy. claudio.chiastra@polimi.it.
Abstract:
The treatment of coronary bifurcations is challenging for interventional cardiologists. The Tryton stent (Tryton Medical, Inc., USA) is one of the few devices specifically designed for coronary bifurcations that underwent large clinical trials. Although the manufacturer provides specific recommendations to position the stent in the bifurcation side branch (SB) according to four radio-opaque markers under angiographic guidance, wrong device positioning may accidentally occur. In this study, the virtual bench testing approach was used to investigate the impact of wrong positioning of the Tryton stent in coronary bifurcations in terms of geometrical and biomechanical criteria. A finite element model of the left anterior descending/first diagonal coronary bifurcation was created with a 45° distal angle and realistic lumen diameters. A validated model of the Tryton stent mounted on stepped delivery balloon was used. All steps of the Tryton deployment sequence were simulated. Three Tryton positions, namely 'proximal', 'recommended', and 'distal' positions, obtained by progressively implanting the stent more distally in the SB, were compared. The 'recommended' case exhibited the lowest ostial area stenosis (44.8 vs. 74.3% ('proximal') and 51.5% ('distal')), the highest diameter at the SB ostium (2.81 vs. 2.70 mm ('proximal') and 2.54 mm ('distal')), low stent malapposition (9.9 vs. 16.3% ('proximal') and 8.5% ('distal')), and the lowest peak wall stress (0.37 vs. 2.20 MPa ('proximal') and 0.71 MPa ('distal')). In conclusion, the study shows that a 'recommended' Tryton stent positioning may be required for optimal clinical results.
Related Concept Videos
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Virtual Work
In static equilibrium, a body can experience an imaginary or virtual movement, such as displacement or rotation. The virtual work done by a force is equal to the dot product of force and virtual displacement in the direction of the force. When it comes to virtually rotating a...
Acute Coronary Syndrome III: Diagnostic Studies
Impact of Groups on Groups
Serum Studies: Renal Function Tests
Coronary Circulation
Coronary circulation begins at the base of the aorta, where two main arteries arise—the left and right coronary arteries. These arteries encircle the heart in the coronary sulcus and supply the...