Measuring Transcutaneous Oxygenation to Validate the Duration Required to Achieve Electrode Equilibration
Nathaniel Chiang1, Jitendra K Jain, Jamie Sleigh
1In the Department of Vascular Surgery, Waikato Hospital, Hamilton, New Zealand, Nathaniel Chiang, MBChB, is a Vascular Fellow; Jamie Sleigh, MBChB, MD, FANZCA, is Consultant Anesthetist; and Thodur Vasudevan, MD, FRACS, is Head of Unit and Consultant Vascular Surgeon. Jitendra K. Jain, MBBS, is Vascular Registrar, The Royal Melbourne Hospital, Melbourne, Australia. Part of this study was presented at the 2013 Annual Scientific Meeting of the Australian and New Zealand Society for Vascular Surgery, Hobart, Tasmania, October 12 to 15, 2015. This study was also given as an oral presentation at the 2017 Symposium of Advanced Wound Care Fall Meeting in Las Vegas, Nevada, October 2017. The authors have disclosed that this research was supported by funding from the University of Auckland, Waikato Medical Research Foundation, Health Research Council, and Royal Australasian College of Surgeons. The authors have disclosed no other financial relationships related to this article. Submitted August 14, 2017; accepted in revised form November 17, 2017.
A minimum of 15 minutes is required for transcutaneous oxygenation measurement (TCOM) electrodes to equilibrate. This duration ensures reliable tissue oxygenation values for assessing wound healing potential.
Area of Science:
- Biomedical Engineering
- Clinical Physiology
- Wound Care
Background:
- Transcutaneous oxygenation measurement (TCOM) systems are vital for assessing tissue viability.
- Current clinical guidelines lack specific recommendations for TCOM electrode equilibration time.
- Establishing a reliable equilibration duration is crucial for accurate tissue oxygenation assessment.
Purpose of the Study:
- To validate the necessary duration for TCOM electrode equilibration in a clinical setting.
- To determine a standardized measurement period for reliable tissue oxygenation values.
Main Methods:
- Minute-by-minute TCOM recordings (TcPO2 and TcPCO2) were collected over 15 minutes.
- Data were gathered from 50 participants across three groups: peripheral vascular disease, no known vascular disease, and healthy volunteers (total 82 limbs).
Main Results:
- Transcutaneous partial pressure of oxygen (TcPO2) decreased, while transcutaneous partial pressure of carbon dioxide (TcPCO2) increased over the 15-minute period.
- Measurements stabilized, reaching equilibrium, with minimal proportional changes between minutes 14 and 15 (TcPO2: 0.8%, TcPCO2: 2.9%).
- TCOM measurement trends were consistent across all participant groups.
Conclusions:
- This study provides the first minute-by-minute analysis of TCOM measurement variations.
- A minimum recording duration of 15 minutes ensures TCOM electrode equilibration for accurate absolute values.
- This validated duration supports reliable assessment of wound healing potential using TCOM.
Related Concept Videos
Oxygen Requirements and Growth Patterns
Reliability and Validity
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Standard Electrode Potentials
Requirements for Human Life
Oxygen
Atmospheric air is only about 20 percent oxygen, but that oxygen is a key component of the chemical reactions that keep the body alive, including the reactions that produce ATP. Brain cells are susceptible to a lack of oxygen because they require a...
Proteins: Dietary Sources and Requirements