Related Experiment Video
Updated: Feb 10, 2026

Encapsulation and Permeability Characteristics of Plasma Polymerized Hollow Particles
Published on: August 16, 2012
Short-Term Deposition of PM2.5 Particles on Contact Lens Surfaces: Effect on Oxygen Permeability and Refractive Index
Zhizhang Dong1, Xiaoyan Ding2, Yong Li3
1b Department of Ophthalmology , The Seconnd affiliated hospital & Yuying Children's hospital of Wenzhou Medical University , Wenzhou , China.
Purposes:
To identify the deposition of fine (≤2.5 μm diameter) particulate matter (PM) particles (PM2.5) on contact lens surfaces and to investigate the effects of such deposition on the oxygen permeability (OP) and refractive index (RI) of contact lenses.
Methods:
A total of 36 contact lenses, including rigid gas permeable (RGP) lens and soft contact lens (SCL), were investigated. RGP lens (n=12) and SCL (n=12) (experimental group) were incubated in a PM2.5 solution for 24 h, after which PM2.5-treated RGP lens (n=6) and SCL (n=6) were further washed for 1 h in phosphate-buffered saline (PBS). All lenses were examined by field emission scanning electron microscopy. OP and RI of all lenses were measured.
Results:
Average-sized PM2.5 particles deposited on RGP contact lens and SCL surfaces after immersion in the PM2.5 solution were 3.192 ± 1.637 and 2.158 ± 1.187/100 μm2, respectively. On RGP lens surfaces, we observed both large (≥2.5 µm diameter) and small (PM2.5) particles. PM2.5 particles were deposited in diffuse patterns, primarily along the honeycomb structural border of SCL, while no PM2.5 particles were found in the honeycomb hole of SCL surfaces. Washing in PBS removed the larger PM particles from RGP lens surfaces, but left copious amounts of PM2.5 particles. In contrast, nearly all PM particles were removed from SCL surfaces after PBS washing. OP values of RGP lens and SCL appeared to be unchanged by PM2.5 deposition. RI values increased in both RGP lens and SCL groups after PM2.5 deposition. However, these increases were not statistically significant, suggesting that PM2.5 deposition itself does not cause fluctuations in contact lens RI.
Conclusions:
Deposition of PM2.5 particles on contact lens surfaces varies according to lens material. PM2.5 particles deposited on SCL, but only large particles on RGP surfaces were able to be removed by washing in PBS and did not appear to alter OP and RI of either lens type.
Related Concept Videos
Short-distance Transport of Resources
Factors Affecting Dissolution: Particle Size and Effective Surface Area
Phase Transitions: Sublimation and Deposition
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Contact-dependent Signaling
Gap Junctions
In animal cells, gap junctions are formed...
Subatomic Particles