Phase Contrast Magnetic Resonance Imaging in the Rat Common Carotid Artery
Published on: September 5, 2018
A Comparison Study of Vector Velocity, Spectral Doppler and Magnetic Resonance of Blood Flow in the Common Carotid
Andreas Hjelm Brandt1, Kristoffer Lindskov Hansen1, Caroline Ewertsen1
1Department of Radiology, Rigshospitalet, Copenhagen University Hospital, Copenhagen, Denmark.
Abstract:
Magnetic resonance phase contrast angiography (MRA) is the gold standard for blood flow evaluation. Spectral Doppler ultrasound (SDU) is the first clinical choice, although the method is angle dependent. Vector flow imaging (VFI) is an angle-independent ultrasound method. The aim of the study was to compare VFI- and SDU-estimated peak systolic velocities (PSV) of the common carotid artery (CCA) with PSV obtained by MRA. Furthermore, intra- and inter-observer agreement was determined. MRA estimates were significantly different from SDU estimates (left CCA: p < 0.001, right CCA: p < 0.001), but not from VFI estimates (left CCA: p = 0.28, right CCA: p = 0.18). VFI measured lower PSV in both CCAs compared with SDU (p < 0.001) with improved precision (VFI: left: 24%, right: 18%; SDU: left 38%, right: 23%). Intra- and inter-observer agreement was almost perfect for VFI and SDU (inter-observer correlation coefficient: VFI 0.88, SDU 0.91; intra-observer correlation coefficient: VFI 0.96, SDU 0.97). VFI is more accurate than SDU in evaluating PSV compared with MRA.
Related Concept Videos
Average and Instantaneous Velocity Vectors
Imaging Studies IV: Magnetic Resonance Imaging
Magnetic Vector Potential
Consider an ideal solenoid with n turns per unit length and radius R. If I is the current through the solenoid, the magnetic field inside the solenoid is expressed as the product of vacuum...
Magnetic Resonance Imaging
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Blood Flow