Evaluation of a multiplex flow immunoassay versus conventional assays in detecting autoantibodies in systemic lupus
1Division of Clinical Immunology, Department of Pathology, Queen Mary Hospital, Pokfulam, Hong Kong.
Introduction:
Conventional diagnostic assays are being replaced with automated multiplex assays, but their performance needs to be evaluated. We compared a multiplex flow immunoassay with conventional techniques in the detection of antinuclear antibodies (ANAs) and antibodies to specific extractable nuclear antigens (ENAs) in serum samples from patients with systemic lupus erythematosus.
Methods:
A total of 140 consecutive Chinese patients with systemic lupus erythematosus and 41 healthy controls were included. The automated BioPlex 2200 ANA Screen assay (Bio-Rad Laboratories, Hercules [CA], US) was compared with indirect immunofluorescence. In addition, use of BioPlex 2200 to detect anti-ENA antibodies was compared with in-house assays of countercurrent immunoelectrophoresis (CIEP), enzyme-linked immunosorbent assay (ELISA), and line blot.
Results:
The sensitivity and specificity of BioPlex in detecting ANAs (91.4% and 95.1%, respectively) were comparable to those of indirect immunofluorescence (90.7% and 85.4%, respectively). Overall, BioPlex achieved the best agreement with ELISA in detecting anti-ENA antibodies: agreement was >90% for most antibody types (κ=0.79-0.94). In contrast, agreement was poorest with CIEP, ranging from 85.6% (κ=0.33) for anti-Sm antibodies to 93.9% (κ=0.88) for anti-Ro antibodies. Overall, BioPlex and ELISA had the highest sensitivity, whereas CIEP had the highest specificity. In terms of disease association, anti-Sm detected by CIEP had the best positive predictive value and specificity for lupus nephritis.
Conclusions:
In a local lupus cohort, BioPlex showed comparable sensitivity to indirect immunofluorescence in detecting ANAs and comparable performance to ELISA in detecting anti-ENA antibodies. However, CIEP was the best method in terms of disease specificity.
Related Concept Videos
Sign Convention
The normal force acts perpendicular to the beam's cross-section and can...
Enolate Mechanism Conventions
Second Order systems II
First Order Systems
When a first-order system is subjected to a unit-step input, its response is characterized by its transfer function. By applying the Laplace transform of the unit-step input to the transfer function, expanding the...
Second Order systems I
By reinterpreting the system, one can derive the closed-loop transfer function, which...
Thermodynamic Systems
Consider an example of tea boiling in a kettle. The...