A Comparison of Three Different Photoscreeners in Children
Purpose:
To compare the results obtained from three non-cycloplegic handheld photorefractometers with cycloplegic autorefractometry (Topcon KR-8100; Topcon Corporation, Tokyo, Japan) measurement in children.
Methods:
The refractive status of 238 eyes in 119 healthy children was assessed. The values acquired using photorefraction with the non-cycloplegic PlusoptiX A12 (Plusoptix GmbH, Nuremberg, Germany), Retinomax K-plus 3 (Righton, Tokyo, Japan), and Spot Vision Screener (Welch Allyn, Skaneateles Falls, NY) devices were compared with those obtained from the cycloplegic Topcon KR-8100. The agreement between the measurements was assessed using the intraclass correlation coefficient.
Results:
The mean age was 10.1 ± 3.2 years (range: 6 to 17 years). The mean spherical value for the right eyes was 0.38 diopters (D) (range: -4.50 to 6.25 D) for the Plusoptix A12; 0.45 D (range: -4.50 to 6.25 D) for the Spot Vision Screener; -1.15 D (range: -8.75 to 6.50 D) for the Retinomax K-plus 3; and 0.62 (range: -4.50 to 6.00) for the Topcon KR-8100. The mean spherical equivalent value for the right eyes was 0.41 D (range: -4.50 to 7.90 D) for the Plusoptix A12; 0.18 D (range: -4.75 to 6.13 D) for the Spot Vision Screener; -1.30 D (range: -10.50 to 6.38 D) for the Retinomax K-plus 3; and 0.67 D (range: -4.00 to 6.00 D) for the Topcon KR-8100 (for the right eyes).
Conclusions:
The photorefractometer method was beneficial in the measurement of refractive errors of school-aged children. The PlusoptiX A12 photorefractometer method may eliminate the need for cycloplegia in the detection of refractive errors in children. [J Pediatr Ophthalmol Strabismus. 2018;55(5):306-311.].
Related Concept Videos
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...
Evolutionary Relationships through Genome Comparisons
Comparison Between Electrical And Gravitational Forces
Since both are inverse square law forces, the distance gets canceled when the ratio of the two forces is considered. Instead, the ratio of the electrical and gravitational forces depends on...
Comparison between RL and RC circuits
Drug Dosing: Infants and Children