Stimulated Motion Suppression (STMS): a New Approach to Break the Resolution Barrier for Ion Trap Mass Spectrometry
Xiaoyu Zhou1, Xinwei Liu2, Spencer Chiang2,3
1State Key Laboratory of Precision Measurement Technology and Instruments, Department of Precision Instrument, Tsinghua University, Beijing, 100084, China. zhouyuzxy@mail.tsinghua.edu.cn.
Abstract:
Ion trap is an excellent platform to perform tandem mass spectrometry (MS/MS), but has an intrinsic drawback in resolving power. Using ion resonant ejection as an example, the resolution degradation can be largely attributed to the broadening of the resonant frequency band (RFB) between ion motion and driving alternative-current (AC). To solve this problem, stimulated motion suppression (STMS) was developed. The key idea of STMS is the use of two suppression alternative-current (SAC) signals, which both have reversed initial phases to the main AC. The SACs can block the unexpected sideband ion resonances (or ejections), therefore playing a key role in sharpening the RFB. The proof-of-concept has been demonstrated through ion trajectory simulations and validated experimentally. STMS provides a new and versatile means for the improvement of the ion trap resolution, which for a long time has reached the bottleneck through conventional methods, e.g., increasing the radio-frequency (RF) voltage and decreasing the mass scan rate. At the end, it is worth noting that the idea of STMS is very general and principally can be applied in any RF device for the purposes of high-resolution mass analysis and ion isolation. Graphical Abstract ᅟ.
Related Concept Videos
High-Resolution Mass Spectrometry (HRMS)
Mass Spectrometry: Overview
Tandem Mass Spectrometry
Mass Spectrometry of Amines
Mass Spectrometry: Isotope Effect
Equation of Motion: Center of Mass
Internal forces between any pair of particles manifest as collinear pairs of equal magnitude but opposite directions,...