Characteristics of a p-Si detector in high energy electron fields
Acta Radiologica. Oncology
|January 1, 1985
Abstract:
Comparison of depth ionization distributions from a silicon semiconductor detector and depth dose curves from a plane parallel ionization chamber show that a semiconductor detector of p-type is well suited for relative electron dosimetry in the energy range of 6 to 20 MeV in Ep,0. Maximum deviations of the order of 1.5 per cent and of 1 mm were obtained down to a phantom depth of about 1 mm. The directional dependence of the detector was about 4 per cent.