Dosimetric characterization of a single crystal diamond detector in X-ray beams for preclinical research
Severin Kampfer1, Nathan Cho2, Stephanie E Combs3
1Department of Radiation Oncology, Technical University of Munich, Klinikum rechts der Isar, Munich, Germany; Physik-Department, Technical University of Munich, Munich, Germany.
Abstract:
The aim of this study was to investigate a single crystal diamond detector, the microDiamond detector from PTW (PTW-Freiburg, Freiburg, Germany), concerning the particular requirements in the set-up and energy range used in small animal radiotherapy (RT) research (around 220kV). We tested it to find out the minimal required pre-irradiation dose, the dose linearity, dose rate dependency and the angular response as well as usability in the small animal radiation research platform, SARRP (Xstrahl Ltd., Camberley, UK). For a stable signal in the range of energies used in the study, we found a required pre-irradiation dose of 10Gy. The dose linearity and dose rate dependence measurements showed a very good performance of the microDiamond detector. Regarding the effect of angular dependency, the variation of the response signal is less than 0.5% within the first 15° of the polar angle. In the azimuthal angle, however, there are differences in detector response up to 20%, depending on the range of energies used in the study. In addition, we compared the detector to a radiosensitive film for a profile measurement of a 5×5mm2 irradiation field. Both methods showed a good accordance with the field size, however, the film has a steeper dose gradient in the penumbra region but also a higher noise than the microDiamond detector. We demonstrated that the microDiamond detector is a useful measurement tool for small animal RT research due to its small size. Nevertheless, it seems to be very important to verify the response of the detector in the given set-up and energy range.
Related Concept Videos
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Ionic Crystal Structures
Most monatomic ions behave as charged spheres, and their attraction for ions of opposite charge is the same in every direction. Consequently, stable structures for ionic compounds result (1) when ions of one charge are surrounded by as many ions as possible of the opposite...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Crystal Growth: Principles of Crystallization
Initiating crystallization involves manipulating the concentration of the solute and the temperature of the solution. Since crystal growth occurs when the ratio of concentration and solubility of the solute in the solvent...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...