Accuracy of SPECT/CT-based lung dose calculation for Holmium-166 hepatic radioembolization before OSEM convergence
Bastiaan J van Nierop1, Jip F Prince1, Rob van Rooij1
1Department of Radiology and Nuclear Medicine, University Medical Centre Utrecht, P.O. Box 85500, Utrecht, GA, 3508, The Netherlands.
Purpose:
In intra-arterial hepatic radioembolization using Holmium-166 (166 Ho) microspheres, a predicted lung-absorbed dose of more than 30 Gy is a contraindication for therapy. Therefore, scout imaging by means of quantitative SPECT of the lungs after a low-dose pretreatment session is essential. Earlier we showed the superiority of Monte Carlo-based iterative SPECT reconstructions over conventional reconstructions due to its quantitative nature, required for dosimetry, at the cost of substantial computation times. In clinical routine, however, the limited available time between scout imaging and therapy constrains its application. To reduce computation times, we investigated the minimum number of iterations required to guarantee a clinical acceptable accuracy in lung dose estimation using patient and phantom data.
Methods:
166 Ho scout SPECT data (range: 222-283 MBq) were used from 10 patients. SPECT images were Monte Carlo-based OSEM reconstructed (effective iterations: 240). Additionally, the 4D XCAT anthropomorphic phantom was used to mimic studies with an injected scout activity of 250 MBq and with varying lung-absorbed doses ranging from 0.9 to 225 Gy for a therapeutic dosage of 15 GBq. These studies were reconstructed in the same way as the patient data, and were also reconstructed using a clinically available, standard OSEM algorithm for comparison. Lung-absorbed dose was determined using VOI analysis as a function of iterations.
Results:
The estimated lung-absorbed dose in nine patients ranged upon MC-based OSEM convergence from 0 to 0.26 Gy for a therapeutic dosage. One patient had an estimated lung absorbed-dose for a therapeutic dosage of 20.3 Gy upon MC-based OSEM convergence, or 18.4 Gy after 40 iterations (-9%). The phantom data showed that the lung-absorbed dose upon OSEM convergence was underestimated by 15% as compared to the actual simulated lung dose, and the dose after 40 iterations was underestimated by 9% as compared to the dose upon convergence. Both underestimations were irrespective of the magnitude of the lung-absorbed dose (0.9 to 225 Gy) and thus can be easily corrected for. The quantitative accuracy of the MC-based OSEM reconstructions (40 iterations, before convergence) outperformed the clinical OSEM reconstruction while estimating the lung dose.
Conclusions:
The number of effective iterations necessary for quantitative estimation of the lung dose using MC-based OSEM can be reduced from 240 to 40. The resulting sixfold reduction in calculation time enables processing of the scout images before therapy administration.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:54Implantation and Monitoring by PET/CT of an Orthotopic Model of Human Pleural Mesothelioma in Athymic Mice
Published on: December 21, 2019
Related Concept Videos
Effect of Hepatic Disease on Pharmacokinetics: Dose Adjustments Due to Hepatic Impairment
Effect of Hepatic Disease on Pharmacokinetics: Drug Dosing and Hepatic Blood Flow
Convergent Evolution
Titration Calculations: Weak Acid - Strong Base
For the titration of 25.00 mL of 0.100 M CH3CO2H with 0.100 M NaOH, the reaction can be represented as:
Titration Calculations: Strong Acid - Strong Base
A titration is carried out for 25.00 mL of 0.100 M HCl (strong acid) with 0.100 M of a strong base NaOH. The pH at different volumes of added base solution can be calculated as follows:
(a) Titrant volume = 0 mL. The solution pH is due to the acid ionization of HCl. Because this is a strong acid, the ionization is complete and the hydronium ion molarity is 0.100 M. The pH of the solution is then:
Improving Translational Accuracy