Possible pseudotransmission of Enterobacter cloacae associated with an endobronchial ultrasound scope
Angela Dickson1, Preeti Kondal1, Lou Hilken1
1Providence Southwest Washington Service Area, Olympia, WA.
Abstract:
During an investigation of positive environmental cultures with Enterobacter cloacae from an endobronchial ultrasound scope, possible pseudotransmission was discovered between 2 patients. All reprocessing steps were adhered to and the original equipment manufacture's quality control assessment of the scope could not determine the root cause. Our findings appear to be the first documented case of pseudotransmission from an endobronchial ultrasound scope and suggest bacterial transmission may exist in endoscopes without an elevator channel.
Insights
This study reports the first suspected pseudotransmission of Enterobacter cloacae via an endobronchial ultrasound scope. It suggests that endoscopes lacking an elevator channel may also pose a risk for bacterial transmission.
Area of Science:
- Infectious Diseases
- Medical Device Contamination
- Microbiology
Background:
- Endobronchial ultrasound (EBUS) scopes are critical for diagnosing lung conditions.
- Proper reprocessing of medical instruments is essential to prevent patient-to-patient transmission of infections.
- Enterobacter cloacae is an opportunistic pathogen that can cause serious infections.
Purpose of the Study:
- To investigate a suspected case of pseudotransmission of Enterobacter cloacae between patients via an EBUS scope.
- To identify the source and mechanism of contamination despite adherence to reprocessing protocols.
- To explore the potential for bacterial transmission in EBUS scopes without elevator channels.
Main Methods:
- Environmental cultures were taken from an EBUS scope.
- Positive cultures for Enterobacter cloacae led to an investigation of patient records and scope handling.
- The original equipment manufacturer's quality control was engaged to assess the scope.
Main Results:
- A possible pseudotransmission event of Enterobacter cloacae was identified between two patients.
- Standard reprocessing procedures did not reveal the source of contamination.
- The scope's quality control assessment could not identify the root cause of the contamination.
Conclusions:
- This represents the first documented case of pseudotransmission linked to an EBUS scope.
- The findings suggest that bacterial transmission may occur even in endoscopes lacking an elevator channel.
- Further research is needed to understand transmission dynamics in duodenoscopes and other endoscopes.
Related Concept Videos
The Scope of Physics
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: