Possible pseudotransmission of Enterobacter cloacae associated with an endobronchial ultrasound scope

Angela Dickson1, Preeti Kondal1, Lou Hilken1

  • 1Providence Southwest Washington Service Area, Olympia, WA.

Insights

This study reports the first suspected pseudotransmission of Enterobacter cloacae via an endobronchial ultrasound scope. It suggests that endoscopes lacking an elevator channel may also pose a risk for bacterial transmission.

Area of Science:

  • Infectious Diseases
  • Medical Device Contamination
  • Microbiology

Background:

  • Endobronchial ultrasound (EBUS) scopes are critical for diagnosing lung conditions.
  • Proper reprocessing of medical instruments is essential to prevent patient-to-patient transmission of infections.
  • Enterobacter cloacae is an opportunistic pathogen that can cause serious infections.

Purpose of the Study:

  • To investigate a suspected case of pseudotransmission of Enterobacter cloacae between patients via an EBUS scope.
  • To identify the source and mechanism of contamination despite adherence to reprocessing protocols.
  • To explore the potential for bacterial transmission in EBUS scopes without elevator channels.

Main Methods:

  • Environmental cultures were taken from an EBUS scope.
  • Positive cultures for Enterobacter cloacae led to an investigation of patient records and scope handling.
  • The original equipment manufacturer's quality control was engaged to assess the scope.

Main Results:

  • A possible pseudotransmission event of Enterobacter cloacae was identified between two patients.
  • Standard reprocessing procedures did not reveal the source of contamination.
  • The scope's quality control assessment could not identify the root cause of the contamination.

Conclusions:

  • This represents the first documented case of pseudotransmission linked to an EBUS scope.
  • The findings suggest that bacterial transmission may occur even in endoscopes lacking an elevator channel.
  • Further research is needed to understand transmission dynamics in duodenoscopes and other endoscopes.

Related Concept Videos

The Scope of Physics01:17

The Scope of Physics

Physics is concerned with the interactions of energy, matter, space, and time, in order to discover the underlying mechanisms that underpin all phenomena. The word "physics" comes from the Greek word "phúsis", which means nature. Physics seeks to comprehend the natural world around us at its most fundamental level. It emphasizes the use of quantitative laws to do this, which could be valuable in other fields that want to push the performance boundaries of present...
53.4K
Ultrasound II: Endoscopic Ultrasound and FibroScan01:25

Ultrasound II: Endoscopic Ultrasound and FibroScan

Endoscopic Ultrasound (EUS) and FibroScan are valuable diagnostic tools in gastroenterology and hepatology, each with specific applications and techniques.
Endoscopic Ultrasound (EUS):
711
Ultrasound I: Abdominal Ultrasonography01:20

Ultrasound I: Abdominal Ultrasonography

Introduction:
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
1.7K
Assessing Blood pressure using a doppler ultrasound01:19

Assessing Blood pressure using a doppler ultrasound

To obtain accurate blood pressure measurements in clinical settings, especially when traditional methods are insufficient, healthcare professionals utilize the Doppler ultrasound technique. This method uses high-frequency sound waves to detect blood flow within the arteries, which is crucial for patients with conditions that complicate circulatory system assessment.
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
2.6K