Related Experiment Video
Updated: Feb 9, 2026

Assessing Early Stage Open-Angle Glaucoma in Patients by Isolated-Check Visual Evoked Potential
Published on: May 25, 2020
Association of Diopsys® Short-duration Transient Visual Evoked Potential Latency with Visual Field Progression in
Richard Trevino1, William E Sponsel2, Carolyn E Majcher1
1Optometrist, Rosenberg School of Optometry, University of the Incarnate Word, San Antonio, Texas, USA.
Aim:
To determine the association of Diopsys® NOVA-LX amplitude and latency abnormality scores with perimetric staging of chronic glaucoma, and to explore potential single-visit short-duration transient visual evoked potential (SD-tVEP) trend detection ability utilizing Humphrey 30-2 field progression data.
Materials And Methods:
Setting: Glaucoma subspecialty clinic. Participants: Treated adult chronic glaucoma patients undergoing SD-tVEP evaluation. Main outcome measures: (1) Proportion of eyes designated as suspect or abnormal by the NOVA-LX multifactorial algorithm were determined as a function of glaucoma severity using the most recent Humphrey visual field analyzer (HVFA) 30-2 field. (2) Association between long-term HVFA-guided progression analysis (GPA) annual slopes and SD-tVEP abnormality was assessed to determine whether a single VEP test might help to identify eyes more prone to progressive visual field (VF) loss.
Results:
One hundred and thirty-three eyes of 84 patients (mean age 68 years) were analyzed. The SD-tVEP abnormality increased proportionately with severity of VF loss under high-contrast (Hc) test conditions for both latency (p = 0.001) and amplitude (p < 0.01). The HVFA progression analysis printouts existed for 91 eyes (mean 12.3 fields per eye/range 5-18). Nearly three-quarters (72.5%) of eyes with mean annual HVFA progression >0.7 dB/year (n = 29) had single-visit VEP latency abnormalities. Fewer than half (46.7%) of the remainder (n = 62) showed latency abnormality. Mean progression for eyes with abnormal vs normal VEP latency was -0.87 ± 0.3 dB/year vs -0.32 ± 0.4 dB/year.
Conclusion:
Diopsys NOVA-LX Hc latency abnormality shows strong association with VF loss among a diverse population of clinical patients undergoing active treatment for chronic glaucoma, and appears likely to afford clinically useful trend-detecting test.
Clinical Significance:
The SD-tVEP has the potential to serve as a single-visit clinical indicator to identify glaucoma patients at high risk for VF progression.How to cite this article: Trevino R, Sponsel WE, Majcher CE, Allen J, Rabin J. Association of Diopsys® Short-duration Transient Visual Evoked Potential Latency with Visual Field Progression in Chronic Glaucoma. J Curr Glaucoma Pract 2018;12(1):29-35.
Related Concept Videos
Visual System
Once through the pupil, the light passes through the lens, a...
Visual Agnosia
Finding Electric Potential From Electric Field
Determining Electric Field From Electric Potential
In general, regardless of whether the electric field is uniform, it points in the direction of decreasing potential because the force on a positive...
Photoreceptors and Visual Pathways
Psychosexual Stages of Personality: Latency
The latency period is not...