Technical Note: Adapting a GE SIGNA PET/MR scanner for radiotherapy
Related Experiment Video
Updated: Feb 9, 2026

Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
Patrik Brynolfsson1, Jan Axelsson1, August Holmberg1
1Department of Radiation Sciences, Umeå University, Umeå, 901 87, Sweden.
Purpose:
Simultaneous collection of PET and MR data for radiotherapy purposes are useful for, for example, target definition and dose escalations. However, a prerequisite for using PET/MR in the radiotherapy workflow is the ability to image the patient in treatment position. The aim of this work was to adapt a GE SIGNA PET/MR scanner to image patients for radiotherapy treatment planning and evaluate the impact on signal-to-noise (SNR) of the MR images, and the accuracy of the PET attenuation correction.
Method:
A flat tabletop and a coil holder were developed to image patients in the treatment position, avoid patient contour deformation, and facilitate attenuation correction of flex coils. Attenuation corrections for the developed hardware and an anterior array flex coil were also measured and implemented to the PET/MR system to minimize PET quantitation errors. The reduction of SNR in the MR images due to the added distance between the coils and the patient was evaluated using a large homogenous saline-doped water phantom, and the activity quantitation errors in PET imaging were evaluated with and without the developed attenuation corrections.
Result:
We showed that the activity quantitation errors in PET imaging were within ±5% when correcting for attenuation of the flat tabletop, coil holder, and flex coil. The SNR of the MRI images were reduced to 74% using the tabletop, and 66% using the tabletop and coil holders.
Conclusion:
We present a tabletop and coil holder for an anterior array coil to be used with a GE SIGNA PET/MR scanner, for scanning patients in the radiotherapy work flow. Implementing attenuation correction of the added hardware from the radiotherapy setup leads to acceptable PET image quantitation. The drop in SNR in MR images may require adjustment of the imaging protocols.
Related Concept Videos
Natural Selection and Adaptation
Beyond physical adaptations,...
Adaptability of Cytoskeletal Filaments
Introduction to Innate and Adaptive Immunity
Innate immunity is the body's natural, nonspecific defense system that acts quickly to protect against pathogens. It incorporates physical barriers like skin and mucous membranes and cellular elements such as phagocytes and natural killer cells. This part of our immune system provides an immediate,...
Adaptive Mechanisms in Cancer Cells
Some of the advantages that cancer cells have on normal cells include - enhanced ability to divide without terminally differentiating, induce new blood vessel formation,...
Adaptations that Reduce Water Loss
Special Features of Adaptive Immunity
The primary cell types involved in adaptive immunity are T cells and B cells. Each type has a unique role in defending the body against pathogens. T cells are responsible for cell-mediated immunity. They identify and eliminate infected cells directly,...
![PET Imaging of Neuroinflammation Using [11C]DPA-713 in a Mouse Model of Ischemic Stroke](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F57243.jpg&w=3840&q=50)