External partial breast irradiation in prone position: how to improve accuracy?
Chris Monten1, Liv Veldeman1, Katrien Vandecasteele1
1a Department of Radiation Oncology , Ghent University Hospital , Belgium.
An indicator clip improves tumor targeting accuracy in accelerated partial breast irradiation (APBI), but a preoperative CT scan in the treatment position offers superior precision. These findings are crucial for optimizing APBI for early-stage breast cancer (ESBC).
Area of Science:
- Radiation Oncology
- Medical Imaging
- Breast Cancer Treatment
Background:
- Accelerated partial breast irradiation (APBI) offers limited incremental benefit over whole breast irradiation (WBI) or no radiotherapy in favorable early-stage breast cancer (ESBC).
- Accurate target coverage and minimal toxicity are paramount for justifying APBI.
- Precise tumor delineation is challenging, particularly in the prone position, with scarce data on accuracy.
Purpose of the Study:
- To evaluate the impact of an indicator clip marking tumor depth on precision and accuracy in APBI.
- To assess the added value of a preoperative CT scan in the treatment position for improving target delineation accuracy.
Main Methods:
- Four radiation oncologists delineated the clinical target volume (CTV) in 12 patients using standard guidelines (CTVstandard).
- Additional delineations included CTVclip (expansion around an indicator clip) and CTVclip_CT (expansion based on clip and preoperative GTV).
- Precision was measured by Jaccard index (JIpairs); accuracy was assessed by the overlap rate between the gross tumor volume (GTV) and CTVs.
Main Results:
- Interobserver precision improved significantly with the indicator clip (JIpairs 0.75) compared to standard delineation (JIpairs 0.38).
- Accuracy, measured as GTV-CTV overlap, improved with the clip (0.67) and was highest with the preoperative CT (0.88).
- Preoperative CT in the treatment position yielded the most optimal high-dose coverage (D95, D90) of the GTV.
Conclusions:
- An indicator clip enhances accurate target coverage for prone APBI but does not fully replace the benefits of a preoperative CT.
- Omitting preoperative CT risks missing the target, especially for small tumors, potentially compromising the APBI concept.
- Increasing target volumes to mitigate missing the target raises questions about the fundamental principles of APBI.
More Related Videos
Related Concept Videos
Improving Translational Accuracy
Improving Translational Accuracy
Uncertainty in Measurement: Accuracy and Precision
Work Done on a System by External Force
In the presence of a non-conservative opposing force, like friction, some part of the work done...
Partial Fractions
Accuracy and Precision