Influence of cortical bone anchorage on the primary stability of dental implants
Anne Kelly de Oliveira Nicolau Mantovani1, Ivete Aparecida de Mattias Sartori1, Luciana Reis Azevedo-Alanis2
1Latin American Institute for Dental Research and Education (ILAPEO), Curitiba, PR, Brazil.
Purpose:
This retrospective chart review study assessed patient records to determine implant insertion torque (IT) and implant stability quotient (ISQ) values during implant placement to evaluate the correlation with cortical bone anchorage (mono- or bicortical).
Methods:
Primary stability data (IT during implant placement surgery and ISQ values immediately after implant placement) and cone beam computed tomography of 33 patients (165 implants) were assessed. Patients were divided into the following groups: G1, implants with apical cortical bone contact; G2, implants with bicortical bone contact (apical and cervical regions); and G3, implants with cervical cortical bone contact.
Results:
Sixty-eight implants were excluded due to cortical bone contact on regions other than implant apical or cervical. Ninety-seven implants were therefore assessed for this study. No implant failure was found after a mean 70.42-month follow-up time. Implants with bicortical anchorage (G2) showed higher IT (64.1 Ncm) during implant placement and higher ISQ values (76) (p < 0.05). Monocortical implants (G1, apical, and G3, cervical) showed similar IT (G1 52.3 and G3 54.3) and ISQ values (G1 71.9 and G3 73) (p > 0.05). No correlation (Pearson correlation coefficient) was found between the two stability measurement devices for the different cortical bone anchorages that were analyzed (G1 0.190, G2 0.039, and G3 - 0.027) (p > 0.05).
Conclusions:
Insertion torque values and implant stability quotients were influenced by cortical bone contact. No significant correlation was found between IT and ISQ values-higher insertion torque values do not necessarily lead to higher implant stability quotients.
Related Concept Videos
Complexation Equilibria: Factors Influencing Stability of Complexes
RNA Stability
Nuclear Stability
To hold positively charged protons together...
Stability
The stability of an LTI system is determined by the roots of its characteristic equation, known as poles. A system is stable if it produces a bounded...
Stability of structures
Pole and System Stability
Simple poles are unique roots of the denominator polynomial. Each simple pole corresponds to a distinct solution to the system's characteristic equation, typically resulting in exponential decay terms in the system's...