Parallel imaging compressed sensing for accelerated imaging and improved signal-to-noise ratio in MRI-based postimplant dosimetry of prostate brachytherapy
Related Experiment Video
Updated: Feb 9, 2026

Author Spotlight: Optimized Lung MRI Protocol with Computationally Efficient Reconstruction Methods
Published on: September 6, 2024
Jeremiah W Sanders1, Hao Song1, Steven J Frank2
1Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, TX.
Purpose:
To investigate the feasibility of using parallel imaging compressed sensing (PICS) to reduce scan time and improve signal-to-noise ratio (SNR) in MRI-based postimplant dosimetry of prostate brachytherapy.
Methods And Materials:
Ten patients underwent low-dose-rate prostate brachytherapy with radioactive seeds stranded with positive magnetic resonance-signal seed markers and were scanned on a Siemens 1.5T Aera. MRI comprised a fully balanced steady-state free precession sequence with two 18-channel external pelvic array coils with and without a rigid two-channel endorectal coil. The fully sampled data sets were retrospectively subsampled with increasing acceleration factors and reconstructed with parallel imaging and compressed sensing algorithms. The images were assessed in a blinded reader study by board-certified care providers. Rating scores were compared for statistically significant differences between reconstruction types.
Results:
Images reconstructed from subsampling up to an acceleration factor of 4 with PICS demonstrated consistently sufficient quality for dosimetry with no apparent loss of SNR, anatomy depiction, or seed/marker conspicuity when compared to the fully sampled images. Images obtained with acceleration factors of 5 or 6 revealed reduced spatial resolution and seed marker contrast. Nevertheless, the reader study revealed that images obtained with an acceleration factor of up to 5 and reconstructed with PICS were adequate-to-good for postimplant dosimetry.
Conclusions:
Combined parallel imaging and compressed sensing can substantially reduce scan time in fully balanced steady-state free precession imaging of the prostate while maintaining adequate-to-good image quality for postimplant dosimetry. The saved scan time can be used for multiple signal averages and improved SNR, potentially obviating the need for an endorectal coil in MRI-based postimplant dosimetry.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Accelerators
The effectiveness of calcium chloride can...
Parallel Resonance
Parallel Processing
Bacterial Signaling