Performance of the Oxylog® 1000 portable ventilator in a hyperbaric environment
Lie Sui An1,2, Loy Shun Ting2, Lee Cheng Chuan3
1Corresponding author: Department of Anaesthesiology, Singapore General Hospital, Outram Road, Singapore 169608. liesuian@gmail.com.
Introduction:
The management of mechanically ventilated patients in the hyperbaric environment requires knowledge of how the physical properties of gases change under pressure and how this affects the operation of the ventilator. The primary objective of this study was to test the performance of the Dräger Oxylog 1000® ventilator in a hyperbaric environment.
Methods:
Each of two ventilators was connected to a mechanical test lung system with an in-built pressure gauge. We used a Wright's respirometer to measure the tidal volumes. The same ventilator settings were tested under varying environmental pressures from ambient (101.3 kPa) to 18 meters' sea water (284 kPa) in a multiplace hyperbaric chamber.
Results:
A decrease was found in tidal volume, decrease in airway pressure and increase in respiratory rate delivered by the Dräger Oxylog 1000 portable ventilator with increasing pressures to 284 kPa.
Discussion:
These findings can be explained by the operating principles of the Oxylog 1000, which is a time-controlled, constant-volume ventilator that functions as a flow chopper. Even between the two Oxylog 1000 ventilators tested there were different absolute changes in tidal volume, airway pressures and respiratory rates at various depths. Hence, the trend of changes in these variables is probably more important than absolute values.
Conclusion:
In summary, understanding the trend of changes in ventilator variables will allow clinicians to make appropriate corrections in ventilator settings and carefully monitor adequacy of ventilation to prevent adverse ventilator-associated events. The Dräger Oxylog 1000 portable ventilator is an adequate back-up ventilator for use with straight-forward, ventilator-dependent patients undergoing hyperbaric treatment.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Mechanical Ventilation I: Indication and Settings