Improved sensitivity and limit-of-detection using a receive-only coil in magnetic particle imaging
Hendrik Paysen1, James Wells, Olaf Kosch
1Physikalisch-Technische Bundesanstalt, Abbestrasse 2-12, 10587 Berlin, Germany.
Abstract:
Magnetic particle imaging (MPI) is an imaging modality capable of quantitatively determining the 3D distribution of a magnetic nanoparticle (MNP) ensemble. In this work, we present a method for reducing the MNP limit of detection by employing a new receive-only coil (Rx-coil) for signal acquisition. The new signal detector is designed to improve the sensitivity and thus quality of reconstructed images. We present characterization measurements conducted with the prototype Rx-coil installed in a preclinical MPI scanner. The gradiometric design of the Rx-coil attenuates the unwanted signal contributions arising from the excitation field, leading to a 17 dB lower background level compared to the conventional dual-purpose coil (TxRx-coil), which is crucial for detecting low amounts of MNP. Network analyzer measurements of the frequency-dependent coil sensitivity, as well as spectral analysis of recorded MPI data demonstrate an overall increase of the coil sensitivity of about +12 dB for the Rx-coil. Comparisons of the sensitivity distributions revealed no significant degradations in terms of homogeneity for the Rx-coil compared to the TxRx-coil in an imaging volume of 6 × 3 × 3 cm3. Finally, the limit of detection was determined experimentally for each coil type using a serial dilution of MNPs, resulting in values of 133 ng of iron for the conventional TxRx-coil and 20 ng for the new Rx-coil, using an acquisition time of 2 s. A linear relationship between the reconstructed signal intensities and the iron mass in the samples was observed with coefficients of determination (R2) of above 99% in the range of the limit of detection to 3 103ng(Fe). These results open the way for improved image quality and faster acquisition time in pre-clinical MPI scanners.
Related Concept Videos
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Motion Of A Charged Particle In A Magnetic Field
The Nucleosome Core Particle
The paradox
Nucleosomes, paradoxically, perform two opposite functions simultaneously. On the one hand, their main responsibility is to protect the delicate DNA strands from physical damage and help achieve a higher compaction ratio. While on the other hand, they must allow polymerase enzymes to access DNA...
Limiting Reactant
Subatomic Particles
Magnetic Resonance Imaging