Related Experiment Video
Updated: Feb 8, 2026

Laryngeal Mask Airway LMA Placement in a Neonatal Patient Simulator Using a Non-Inflatable Supraglottic Airway SGA
Published on: July 14, 2023
Direct and indirect low skill fibre-optic intubation: A randomised crossover manikin study of six supraglottic airway
Sau Yee Chow1, Yan Ru Tan1, Theodore Gar Ling Wong1
1Department of Anaesthesiology, Singapore General Hospital, Outram Road, Singapore.
Background And Aims:
Fibre-optic intubation (FOI) through supraglottic airway devices (SADs) is useful in the management of the difficult airway. We compared two methods of FOI through seven SADs in a randomised crossover manikin study to assess each device's performance and discuss implications on SAD selection.
Methods:
Thirty anaesthetsiologists, 15 seniors and 15 juniors, each performed low skill FOI (LSFOI) with seven SADs using both 'direct' and 'indirect' methods. The order of method and device used were randomised. The primary end point was success rate of intubation; secondary end points were time taken for intubation, incidence of difficulties with direct and indirect LSFOI and operator device preference. Statistical analysis was with univariable analysis and comparison of proportions.
Results:
Data from six devices were analysed due to a protocol breach with one SAD. There was no difference in intubation success rate across all SADs and intubation methods. Intubation time was significantly shorter in AmbuAuragain than other SADs and shorter with the direct method of LSFOI than the indirect method (mean difference of 6.9 s, P = 0.027). Ambu Auragain had the least SAD and bronchoscope-related difficulties. Seniors had significantly shorter mean intubation times than juniors by 11.6 s (P = 0.0392). The most preferred SAD for both methods was AmbuAuragain.
Conclusion:
Low skill FOI consistently achieves a high intubation success rate regardless of experience, choice of method, or SAD used. SAD design features may significantly affect the performance of low skill FOI.
Related Concept Videos
Crossover Experiments
Crossover designs are performed even with smaller sample sizes since the samples can act as their controls. These are better than simple randomized trials since patients are exposed to all the treatments.
Indirect Motor Pathways
The vestibulospinal tract originates in the vestibular nuclei of the brainstem. The vestibular system detects changes in...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
EDTA: Indirect and Alkalimetric Titration
Suctioning the Nasopharyngeal Airway
Equipment Required
Suctioning the Oropharyngeal Airway
After assembling the equipment, the nurse should practice hand hygiene and don appropriate PPE according to infection control guidelines to avoid the...