Comparison of three non-invasive hemodynamic monitoring methods in critically ill children
Chanapai Chaiyakulsil1, Marut Chantra1, Poomiporn Katanyuwong2
1Department of Pediatrics, Division of Pediatric Critical Care, Department of Pediatric, Faculty of Medicine, Ramathibodi Hospital, Mahidol University, Bangkok, Thailand.
Introduction:
Hemodynamic parameters measurements were widely conducted using pulmonary artery catheter (PAC) with thermodilution as a reference standard. Due to its technical difficulties in children, transthoracic echocardiography (TTE) has been widely employed instead. Nonetheless, TTE requires expertise and is time-consuming. Noninvasive cardiac output monitoring such as ultrasonic cardiac output monitor (USCOM) and electrical velocimetry (EV) can be performed rapidly with less expertise requirement. Presently, there are inconsistent evidences, variable precision, and reproducibility of EV, USCOM and TTE measurements. Our objective was to compare USCOM, EV and TTE in hemodynamic measurements in critically ill children.
Materials And Methods:
This was a single center, prospective observational study in critically ill children. Children with congenital heart diseases and unstable hemodynamics were excluded. Simultaneous measurements of hemodynamic parameters were conducted using USCOM, EV, and TTE. Inter-rater reliability was determined. Bland-Altman plots were used to analyse agreement of assessed parameters.
Results:
Analysis was performed in 121 patients with mean age of 4.9 years old and 56.2% of male population. Interrater reliability showed acceptable agreement in all measured parameters (stroke volume (SV), cardiac output (CO), velocity time integral (VTI), inotropy (INO), flow time corrected (FTC), aortic valve diameter (AV), systemic vascular resistance (SVR), and stroke volume variation (SVV); (Cronbach's alpha 0.76-0.98). Percentages of error in all parameters were acceptable by Bland-Altman analysis (9.2-28.8%) except SVR (30.8%) and SVV (257.1%).
Conclusion:
Three noninvasive methods might be used interchangeably in pediatric critical care settings with stable hemodynamics. Interpretation of SVV and SVR measurements must be done with prudence.
Related Concept Videos
Critical Numbers and the Closed Interval Method
Renal Drug Clearance: Comparison Between Renal Excretion Methods
Renal clearance is often associated with the renal glomerular filtration rate (GFR), which represents the rate at which plasma is filtered through the glomeruli in the kidney. When drug reabsorption is minimal and there is no active secretion, renal clearance is closely related to the...
Critical Region, Critical Values and Significance Level
In hypothesis testing, a sample statistic is converted to a test statistic using z, t, or chi-square distribution. A critical region is an area under the curve in probability distributions demarcated by the critical value. When the test statistic falls in this region, it suggests that the null hypothesis must be rejected. As this region contains all those values of the...
Critical Values
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Critical Thinking I