Instantaneous wave-free ratio and fractional flow reserve in clinical practice
R Pisters1,2,3, M Ilhan4,5, L F Veenstra4,5
1Department of Cardiology, Zuyderland Medical Centre, Heerlen, The Netherlands. rpisters@rijnstate.nl.
Objectives:
To compare fractional flow reserve (FFR) and instantaneous wave-free ratio (iFR) measurements in an all-comer patient population with moderate coronary artery stenoses.
Background:
Visual assessment of the severity of coronary artery stenoses is often discordant in moderate lesions. FFR allows reliable functional severity assessment in these cases but requires adenosine-induced hyperaemia with associated additional time, costs and side effects. The iFR is a hyperaemia-independent index.
Methods And Results:
Between November 2015 and February 2017, 356 consecutive patients were included in whom 515 coronary stenoses were measured using both iFR and FFR. Mean iFR and FFR were 0.90 ± 0.09 and 0.86 ± 0.08, respectively. iFR correlated well with FFR [r = 0.75; p < 0.001]. Receiver operating characteristic analysis identified an area under the curve of 0.92. An iFR-only strategy with a treatment cut-off ≤0.89 revealed a diagnostic classification agreement with the FFR-only strategy in 420 lesions (82%) with a sensitivity of 87%, a specificity of 80%, a positive predictive value of 56% and a negative predictive value of 96%.
Conclusions:
Real-time iFR measurements have good negative predictive value compared to FFR, but moderate diagnostic accuracy (82%). It exposes fewer patients to adenosine, reduces procedure time and costs. Further prospective trials are needed to evaluate specific clinical settings, cut-off values and endpoints.
More Related Videos
06:39Ultrasound Based Assessment of Coronary Artery Flow and Coronary Flow Reserve Using the Pressure Overload Model in Mice
Published on: April 13, 2015
05:07Author Spotlight: Improved Localization and Monitoring of Coronary Flow Reserve Using Modified PLAX View in Mice
Published on: August 25, 2023
Related Concept Videos
Instantaneous Velocity - II
Instantaneous Acceleration
Instantaneous Power
Instantaneous Velocity - I
Instantaneous Center of Zero Velocity
To analyze this, consider two points on the wheel: point A and point B. The absolute velocity of point B can be expressed as the vector sum of the absolute velocity of point A and the relative velocity of point B with respect to point A. To simplify this analysis,...
The Wave Nature of Light