Impact of strut dimensions and vessel caliber on thrombosis risk of bioresorbable scaffolds using hemodynamic metrics
Michael Stiehm1, Carolin Wüstenhagen1, Stefan Siewert1
1Institute for ImplantatTechnology and Biomaterials e.V., Friedrich-Barnewitz-Str. 4, 18119 Rostock-Warnemünde, Germany.
Insights
Bioresorbable scaffolds (BRS) may increase thrombosis risk in small coronary vessels. Thicker struts and smaller vessel sizes elevate platelet activation, transport, and aggregation, suggesting bulky stents in narrow arteries pose higher risks.
Area of Science:
- Cardiovascular Engineering
- Biomedical Engineering
- Computational Fluid Dynamics
Background:
- Bioresorbable scaffolds (BRS) are promising for treating stenosed coronary vessels.
- Current clinical data indicate a higher-than-expected thrombosis risk associated with BRS.
- Understanding the hemodynamic factors influencing thrombosis is crucial for BRS development.
Purpose of the Study:
- To evaluate the thrombosis risk of coronary stents/scaffolds using computational fluid dynamics (CFD).
- To introduce and utilize three hemodynamic metrics: platelet shear number (PSN), platelet convection number (PCN), and platelet aggregation number (PAN).
- To compare the thrombosis risk between thick-strut and thin-strut BRS designs in small and medium coronary vessels.
Main Methods:
- Transient computational fluid dynamics (CFD) simulations were employed.
- Analysis focused on platelet activation and diffusion (PSN), convective transport (PCN), and aggregation (PAN).
- Two stent designs (thick-strut vs. thin-strut) were simulated in two vessel sizes (RVD 2.25 mm vs. 2.70 mm).
Main Results:
- Thick-strut designs exhibited higher PSN, PCN, and PAN compared to thin-strut designs in both vessel sizes.
- Smaller vessel size (RVD 2.25 mm) significantly increased PSN and PCN for both stent designs.
- Sensitivity analysis indicated PSN is primarily influenced by vessel size, while PCN and PAN are sensitive to strut thickness.
Conclusions:
- Bulky stents implanted in small-caliber coronary arteries may substantially increase thrombosis risk.
- Strut thickness is a critical device-related factor influencing platelet transport and aggregation.
- Vessel size is a key lesion-related factor affecting platelet activation and diffusion.
Abstract:
Bioresorbable scaffolds (BRS) promise to be the treatment of choice for stenosed coronary vessels. But higher thrombosis risk found in current clinical studies limits the expectations. Three hemodynamic metrics are introduced to evaluate the thrombosis risk of coronary stents/scaffolds using transient computational fluid dynamics (CFD). The principal phenomena are platelet activation and effective diffusion (platelet shear number, PSN), convective platelet transport (platelet convection number, PCN) and platelet aggregation (platelet aggregation number, PAN) were taken into consideration. In the present study, two different stent designs (thick-strut vs. thin-strut design) positioned in small- and medium-sized vessels (reference vessel diameter, RVD=2.25 mm vs. 2.70 mm) were analyzed. In both vessel models, the thick-strut design induced higher PSN, PCN and PAN values than the thin-strut design (thick-strut vs. thin-strut: PSN=2.92/2.19 and 0.54/0.30; PCN=3.14/1.15 and 2.08/0.43; PAN: 14.76/8.19 and 20.03/10.18 for RVD=2.25 mm and 2.70 mm). PSN and PCN are increased by the reduction of the vessel size (PSN: RVD=2.25 mm vs. 2.70 mm=5.41 and 7.30; PCN: RVD=2.25 mm vs. 2.70 mm=1.51 and 2.67 for thick-strut and thin-strut designs). The results suggest that bulky stents implanted in small caliber vessels may substantially increase the thrombosis risk. Moreover, sensitivity analyses imply that PSN is mostly influenced by vessel size (lesion-related factor), whereas PCN and PAN sensitively respond to strut-thickness (device-related factor).
Related Concept Videos
Venous Thrombosis I: Introduction
Support Reactions in Three Dimensions
Ball and Socket Joint is one of the supports allowing free rotation about any axis. This freedom of rotation is...
Relative Velocity in One Dimension
Relative Risk
Impact of Groups on Groups
Relative Velocity in Two Dimensions