Aqueous volume loss associated with occlusion break surge in phacoemulsifiers from 4 different manufacturers
Carolina Aravena1, David W Dyk1, Andrew Thorne1
1From the Stein Eye Institute and Department of Ophthalmology (Aravena, Thorne, Miller), David Geffen School of Medicine, University of California Los Angeles, Los Angeles and Alcon Research Ltd. (Dyk, Fanney), Lake Forest, California, USA Dr. Aravena is now with the Departamento de Oftalmología, Escuela de Medicina, Pontificia Universidad Católica de Chile, Santiago, Chile.
Purpose:
To evaluate aqueous volume losses associated with occlusion breaks at varying vacuum limits in phacoemulsification systems from 4 different manufacturers.
Setting:
Alcon Research Ltd., Lake Forest, California, USA.
Design:
Experimental study.
Methods:
The anterior chamber was modeled using the spring eye model. Systems tested included the Centurion, Whitestar Signature, Stellaris PC, and EVA. Occlusion breaks were actuated at vacuum limits of 200, 300, 400, 500, and 600 mm Hg and a target intraocular pressure of 55 mm Hg.
Results:
The model eye piston reached its measurement limit just below 600 mm Hg on the EVA and just above 400 mm Hg on the Stellaris PC. Higher vacuum limits could not be tested on these 2 units. Surge volumes varied from 17 to 77 μL on the Centurion, 30 to 103 μL on the Whitestar Signature, 67 μL to 163 μL on the Stellaris PC, and 47 to 165 μL on the EVA. Assuming an average phakic eye aqueous volume of 250 μL, these μL values correspond to percent aqueous volume losses of 7% to 31% on the Centurion, 12% to 41% on the Whitestar Signature, 27% to 65% on the Stellaris PC, and 19% to 66% on the EVA. Surge responses increased on all machines with each increment in vacuum limit. The Centurion had the lowest surge volumes across all vacuum limits.
Conclusions:
Occlusion break surge volumes vary considerably across phacoemulsification platforms. Severe chamber shallowing might occur if an occlusion break occurs under high vacuum on some systems.
More Related Videos
05:38Micromechanical Tension Testing of Additively Manufactured 17-4 PH Stainless Steel Specimens
Published on: April 7, 2021
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024
Related Concept Videos
Aqueous Solutions and Heats of Hydration
When ionic compounds dissolve in water, the ions in the solid separate and disperse uniformly throughout the solution because water molecules surround and solvate the ions, reducing the strong electrostatic forces between them. This process...
Chemical Reactions in Aqueous Solutions
Fixing Double-strand Breaks
Fixing Double-strand Breaks
Line Loss
Line loss impacts power delivery efficiency in a balanced three-phase circuit. The symmetry in such a circuit simplifies the...
Steel Manufacturing
During this smelting process, limestone plays a crucial role by forming slag. Slag captures impurities within the molten iron, such...