A Zátonyi1,2,3, F Fedor2,3,4, Zs Borhegyi4
1Research Group for Implantable Microsystems, Faculty of Information Technology and Bionics, Pázmány Péter Catholic Uiversity, Budapest, Hungary.
Objective:
Intracranial EEG (iEEG) or micro-electrocorticography (µECoG) microelectrodes offer high spatial resolution in recordings of neuronal activity from the exposed brain surface. Reliability of dielectric substrates and conductive materials of these devices are under intensive research in terms of functional stability in biological environments.
Approach:
The aim of our study is to investigate the stability of electroplated platinum recording sites on 16-channel, 8 micron thick, polyimide based, flexible µECoG arrays implanted underneath the skull of rats. Scanning electron microscopy and electrochemical impedance spectroscopy was used to reveal changes in either surface morphology or interfacial characteristics. The effect of improved surface area (roughness factor = 23 ± 0.12) on in vivo recording capability was characterized in both acute and chronic experiments.
Main Results:
Besides the expected reduction in thermal noise and enhancement in signal-to-noise ratio (up to 39.8), a slight increase in the electrical impedance of individual sites was observed, as a result of changes in the measured interfacial capacitance. In this paper, we also present technology processes and protocols in detail to use such implants without crack formation of the porous platinum surfaces.
Significance:
Our findings imply that black-platinum coating deposited on the recording sites of flexible microelectrodes (20 microns in diameter) provides a stable interface between tissue and device.
More Related Videos
Related Concept Videos
Polymers
Polymers
Detection of Black Holes
Their closest cousins are neutron stars, which are composed almost entirely of neutrons packed against each other, making them extremely dense. A neutron star has the same mass as the Sun but its diameter is only a few kilometers. Therefore, the escape velocity from their surface is close to the speed of light.
Not until the 1960s, when the first neutron...
Nuclear Stability
To hold positively charged protons together...
RNA Stability
Equivalence: In Vitro and In Vivo Bioequivalence